{"paper":{"title":"A one-dimensional model for aspiration therapy in blood vessels","license":"http://creativecommons.org/licenses/by-nc-sa/4.0/","headline":"","cross_cats":["cs.NA"],"primary_cat":"math.NA","authors_text":"Michael Herty, Michael Neidlin, Niklas Kolbe","submitted_at":"2024-03-08T18:04:56Z","abstract_excerpt":"Aspiration thrombectomy is a treatment option for ischemic stroke due to occlusions in large vessels. During the therapy a device is inserted into the vessel and suction is applied. A new one-dimensional model is introduced that is capable of simulating this procedure while accounting for the fluid-structure interactions in blood flow. To solve the coupling problem at the tip of the device a problem-suited Riemann solver is constructed based on relaxation of the hyperbolic model. Numerical experiments investigating the role of the catheter size and the suction forces are presented."},"claims":{"count":0,"items":[],"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"source":{"id":"2403.05494","kind":"arxiv","version":1},"verdict":{"id":null,"model_set":{},"created_at":null,"strongest_claim":"","one_line_summary":"","pipeline_version":null,"weakest_assumption":"","pith_extraction_headline":""},"integrity":{"clean":true,"summary":{"advisory":0,"critical":0,"by_detector":{},"informational":0},"endpoint":"/pith/2403.05494/integrity.json","findings":[],"available":true,"detectors_run":[],"snapshot_sha256":"c28c3603d3b5d939e8dc4c7e95fa8dfce3d595e45f758748cecf8e644a296938"},"references":{"count":0,"sample":[],"resolved_work":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57","internal_anchors":0},"formal_canon":{"evidence_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"author_claims":{"count":0,"strong_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"builder_version":"pith-number-builder-2026-05-17-v1"}