{"paper":{"title":"A Novel Method of Encoded Multiplexing Readout for Micro-pattern Gas Detectors","license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","headline":"","cross_cats":["hep-ex"],"primary_cat":"physics.ins-det","authors_text":"Bin-Xiang Qi, Heng Ji, Hong-Bang Liu, Qi An, Shu-Bin Liu, Si-Yuan Ma, Wen-Qian Huang, Zhong-Tao Shen","submitted_at":"2015-09-07T23:47:01Z","abstract_excerpt":"The requirement of a large number of electronic channels poses a big challenge for Micro-pattern Gas Detector (MPGD) to achieve good spatial resolution. By using the redundancy that at least two neighboring strips record the signal of a particle, a novel method of encoded multiplexing readout for MPGDs is presented in this paper. The method offers a feasible and easily-extensible way of encoding and decoding, and can significantly reduce the number of readout channels. A verification test was carried out on a 5*5 cm2 Thick Gas Electron Multiplier (THGEM) detector using a 8 keV Cu X-ray source "},"claims":{"count":0,"items":[],"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"source":{"id":"1509.02229","kind":"arxiv","version":1},"verdict":{"id":null,"model_set":{},"created_at":null,"strongest_claim":"","one_line_summary":"","pipeline_version":null,"weakest_assumption":"","pith_extraction_headline":""},"references":{"count":0,"sample":[],"resolved_work":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57","internal_anchors":0},"formal_canon":{"evidence_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"author_claims":{"count":0,"strong_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"builder_version":"pith-number-builder-2026-05-17-v1"}