{"paper":{"title":"A 83Krm Source for Use in Low-background Liquid Xenon Time Projection Chambers","license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","headline":"","cross_cats":[],"primary_cat":"physics.ins-det","authors_text":"A. Manzur, D. N. McKinsey, L. W. Kastens, S. B. Cahn, S. Bedikian","submitted_at":"2009-12-11T20:44:40Z","abstract_excerpt":"We report the testing of a charcoal-based Kr-83m source for use in calibrating a low background two-phase liquid xenon detector. Kr-83m atoms produced through the decay of Rb-83 are introduced into a xenon detector by flowing xenon gas past the Rb-83 source. 9.4 keV and 32.1 keV transitions from decaying 83Krm nuclei are detected through liquid xenon scintillation and ionization. The characteristics of the Kr-83m source are analyzed and shown to be appropriate for a low background liquid xenon detector. Introduction of Kr-83m allows for quick, periodic calibration of low background noble liqui"},"claims":{"count":0,"items":[],"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"source":{"id":"0912.2337","kind":"arxiv","version":3},"verdict":{"id":null,"model_set":{},"created_at":null,"strongest_claim":"","one_line_summary":"","pipeline_version":null,"weakest_assumption":"","pith_extraction_headline":""},"references":{"count":0,"sample":[],"resolved_work":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57","internal_anchors":0},"formal_canon":{"evidence_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"author_claims":{"count":0,"strong_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"builder_version":"pith-number-builder-2026-05-17-v1"}