{"paper":{"title":"Validation of a New Method for Stroke Volume Variation Assessment: a Comparaison with the PiCCO Technique","license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","headline":"","cross_cats":["math-ph","math.MP"],"primary_cat":"physics.med-ph","authors_text":"Andry Van De Louw, Claire M\\'edigue (INRIA Rocquencourt), Fran\\c{c}ois Cottin (UBIAE), Taous-Meriem Laleg-Kirati (INRIA Bordeaux - Sud-Ouest), Yves Papelier","submitted_at":"2009-11-04T14:40:43Z","abstract_excerpt":"This paper proposes a novel, simple and minimally invasive method for stroke volume variation assessment using arterial blood pressure measurements. The arterial blood pressure signal is reconstructed using a semi-classical signal analysis method allowing the computation of a parameter, called the first systolic invariant INVS1. We show that INVS1 is linearly related to stroke volume. To validate this approach, a statistical comparaison between INVS1 and stroke volume measured with the PiCCO technique was performed during a 15-mn recording in 21 mechanically ventilated patients in intensive ca"},"claims":{"count":0,"items":[],"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"source":{"id":"0911.0837","kind":"arxiv","version":3},"verdict":{"id":null,"model_set":{},"created_at":null,"strongest_claim":"","one_line_summary":"","pipeline_version":null,"weakest_assumption":"","pith_extraction_headline":""},"references":{"count":0,"sample":[],"resolved_work":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57","internal_anchors":0},"formal_canon":{"evidence_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"author_claims":{"count":0,"strong_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"builder_version":"pith-number-builder-2026-05-17-v1"}