{"paper":{"title":"Development of a 3-D Position Sensitive Neutron Detector Based on Organic Scintillators with Double Side SiPM Readout","license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","headline":"","cross_cats":["nucl-ex"],"primary_cat":"physics.ins-det","authors_text":"Yang Tian, Yidong Fu, Yuanjing Li, Yulan Li","submitted_at":"2018-10-31T17:31:35Z","abstract_excerpt":"A 3-D position sensitive neutron detector is being developed based on a plastic scintillator array. A double side SiPM readout is used to determine the depth of interaction (DOI) in each scintillator unit. In the preliminary test, the DOI in a 254 x 6 x 6 mm3 SP101 plastic scintillator is measured at different positions using a collimated Co-60 source. The SiPM (KETEK PM6660) signals are recorded by a 2.5 GS/s digital oscilloscope. The DOI results are calculated using both the amplitude and the temporal information. Position resolutions (FWHM) of 2.5 cm and 6.6 cm are realized, respectively. A"},"claims":{"count":0,"items":[],"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"source":{"id":"1810.13420","kind":"arxiv","version":1},"verdict":{"id":null,"model_set":{},"created_at":null,"strongest_claim":"","one_line_summary":"","pipeline_version":null,"weakest_assumption":"","pith_extraction_headline":""},"references":{"count":0,"sample":[],"resolved_work":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57","internal_anchors":0},"formal_canon":{"evidence_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"author_claims":{"count":0,"strong_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"builder_version":"pith-number-builder-2026-05-17-v1"}