{"paper":{"title":"Conditioning the gamma spectrometer for activity measurement at very high background","license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","headline":"","cross_cats":["nucl-ex"],"primary_cat":"physics.ins-det","authors_text":"Liguo Zhang, Weihua Yan, Zhao Zhang, Zhigang Xiao","submitted_at":"2013-05-04T09:49:44Z","abstract_excerpt":"The application of a high purity germanium (HPGe) gamma spectrometer in determining the fuel element burnup in a future reactor is studied. The HPGe detector is exposed by a Co60 source with varying irradiation rate from 10 kcps to 150 kcps to simulate the input counting rate in real reactor environment. A Cs137 and a Eu152 source are positioned at given distances to generate certain event rate in the detector with the former being proposed as a labeling nuclide to measure the burnup of fuel element. It is shown that both the energy resolution slightly increasing with the irradiation rate and "},"claims":{"count":0,"items":[],"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"source":{"id":"1305.0903","kind":"arxiv","version":1},"verdict":{"id":null,"model_set":{},"created_at":null,"strongest_claim":"","one_line_summary":"","pipeline_version":null,"weakest_assumption":"","pith_extraction_headline":""},"references":{"count":0,"sample":[],"resolved_work":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57","internal_anchors":0},"formal_canon":{"evidence_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"author_claims":{"count":0,"strong_count":0,"snapshot_sha256":"258153158e38e3291e3d48162225fcdb2d5a3ed65a07baac614ab91432fd4f57"},"builder_version":"pith-number-builder-2026-05-17-v1"}