{"arxiv_id":"2611.17187","title":"The Gator low-background spectrometer: readout upgrade and photomultiplier tube radioassay for XLZD","abstract":"We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay. The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.","primary_cat":"physics.ins-det","cross_cats":["nucl-ex"],"submitted_at":"2026-11-12T17:52:55+00:00","latest_version":1,"source":"arxiv","source_label":"arXiv","source_url":null,"has_pdf":true,"has_source":false,"verdict":"ACCEPT","confidence":"HIGH","one_line_summary":"The Gator spectrometer's readout upgrade improves energy resolution by 17% at 2614.5 keV and suppresses low-energy noise; radioassay of 16 Hamamatsu R17317 PMTs with improved stems shows reduced 238U and 232Th activities by a factor of three.","novelty_score":4.0,"correctness_risk":"low","formal_verification":"none","upvotes":0,"downvotes":0,"pith_score":0,"inbound_count":0,"_links":{"html":"https://pith.science/paper/2611.17187","verdict":"/api/v1/verdicts/2611.17187"}}