Evidence of delayed light emission of TetraPhenyl Butadiene excited by liquid Argon scintillation light

classification ⚛️ physics.ins-det nucl-ex
keywords photonsbutadienescintillationtetraphenyltimeargonlightliquid
0
0 comments X
read the original abstract

TetraPhenyl Butadiene is the wavelength shifter most widely used in combination with liquid Argon. The latter emits scintillation photons with a wavelength of 127 nm that need to be downshifted to be detected by photomultipliers with glass or quartz windows. TetraPhenyl Butadiene has been demonstrated to have an extremely high conversion efficiency, possibly higher than 100 % for 127 nm photons, while there is no precise information about the time dependence of its emission. It is usually assumed to be exponentially decaying with a characteristic time of the order of one ns, as an extrapolation from measurements with exciting radiation in the near UV. This work shows that TetraPhenyl Butadiene, when excited by 127 nm photons, reemits photons not only with a very short decay time, but also with slower ones due to triplet states de-excitations. This fact can strongly contribute to clarify the anomalies in liquid Argon scintillation light reported in literature since seventies, namely the inconsistency in the measured values of the long decay time constant and the appearance of an intermediate component. Similar effects should be also expected when the TPB is used in combination with Helium and Neon, that emit scintillation photons with wavelengths shorter than 127 nm.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.

Forward citations

Cited by 1 Pith paper

Reviewed papers in the Pith corpus that reference this work. Sorted by Pith novelty score.

  1. Procurement and Purification of Liquid Argon for the LEGEND-200 Experiment

    physics.ins-det 2026-05 unverdicted novelty 3.0

    A dedicated purification system using copper catalyst and molecular sieves improved liquid argon purity for LEGEND-200 from τ_t ≈ 0.9 μs to 1.3 μs, with final measured value of 1.16 μs limited by accidental nitrogen c...