Retrospective analysis of 302 pediatric CIED surgeries yields weight thresholds (epicardial-only below 21 kg, dual use 21-66 kg, endocardial above) and lead-length constraints to guide realistic MRI heating simulations.
Impact of imaging landmark on the risk of MRI- related heating near implanted medical devices like cardiac pacemaker leads,
1 Pith paper cite this work, alongside 48 external citations. Polarity classification is still indexing.
1
Pith paper citing it
48
external citations · OpenAlex
fields
physics.med-ph 1years
2026 1verdicts
UNVERDICTED 1representative citing papers
citing papers explorer
-
Weight-Guided Constraints for Body Model and Lead Selection in Pediatric CIED MRI Safety Simulations
Retrospective analysis of 302 pediatric CIED surgeries yields weight thresholds (epicardial-only below 21 kg, dual use 21-66 kg, endocardial above) and lead-length constraints to guide realistic MRI heating simulations.