A bolometer made of superfluid helium-3, instrumented with 400 nm nanowire resonators and SQUID readout, was operated at 0.3 mK, with two-wire heat calibration and frequency multiplexing demonstrated.
Using vibrating wire in non-linear regime as a thermometer in superfluid $^3$He-B
1 Pith paper cite this work. Polarity classification is still indexing.
abstract
Vibrating wires are common temperature probes in $^3$He experiments. By measuring mechanical resonance of a wire driven by AC current in magnetic field one can directly obtain temperature-dependent viscous damping. This is easy to do in a linear regime where wire velocity is small enough and damping force is proportional to velocity. At lowest temperatures in superfluid $^3$He-B a strong non-linear damping appears and linear regime shrinks to a very small velocity range. Expanding measurements to the non-linear area can significantly improve sensitivity. In this note I describe some technical details useful for analyzing such temperature measurements.
citation-role summary
citation-polarity summary
fields
physics.ins-det 1years
2025 1verdicts
CONDITIONAL 1roles
background 1polarities
unclear 1representative citing papers
citing papers explorer
-
Development of superfluid helium-3 bolometry using nanowire resonators with SQUID readout for the QUEST-DMC experiment
A bolometer made of superfluid helium-3, instrumented with 400 nm nanowire resonators and SQUID readout, was operated at 0.3 mK, with two-wire heat calibration and frequency multiplexing demonstrated.