A new field validation framework for wearable light loggers demonstrates high agreement (R²=0.988-0.990) with reference instruments but systematic underestimation, with performance stabilizing at ~100 diverse scenes and consistent across sites.
Comparison and correction of the light sensor output from 48 wearable light exposure devices by using a side-by-side field calibration method
1 Pith paper cite this work. Polarity classification is still indexing.
1
Pith paper citing it
fields
physics.ins-det 1years
2026 1verdicts
CONDITIONAL 1representative citing papers
citing papers explorer
-
Scene-based field validation of wearable light loggers
A new field validation framework for wearable light loggers demonstrates high agreement (R²=0.988-0.990) with reference instruments but systematic underestimation, with performance stabilizing at ~100 diverse scenes and consistent across sites.