EXO-200 measured the absolute ionization and scintillation yields of liquid xenon at 1.1-2.6 MeV, finding a W-value of 11.5 eV and a recombination-independent energy scale about 15% below the NEST simulation prediction.
MiX: A Position Sensitive Dual-Phase Liquid Xenon Detector
1 Pith paper cite this work. Polarity classification is still indexing.
abstract
The need for precise characterization of dual-phase xenon detectors has grown as the technology has matured into a state of high efficacy for rare event searches. The Michigan Xenon detector was constructed to study the microphysics of particle interactions in liquid xenon across a large energy range in an effort to probe aspects of radiation detection in liquid xenon. We report the design and performance of a small 3D position sensitive dual-phase liquid xenon time projection chamber with high light yield ($L_y^{122}=15.2 $pe/keV at zero field), long electron lifetime ($\tau > 200 \mu$s), and excellent energy resolution ($\sigma/E = 1\%$ for 1,333 keV gamma rays in a drift field of 200 V/cm). Liquid xenon time projection chambers with such high energy resolution may find applications not only in dark matter direct detection searches, but also in neutrinoless double beta decay experiments and other applications.
fields
physics.ins-det 1years
2019 1verdicts
ACCEPT 1representative citing papers
citing papers explorer
-
Measurement of the scintillation and ionization response of liquid xenon at MeV energies in the EXO-200 experiment
EXO-200 measured the absolute ionization and scintillation yields of liquid xenon at 1.1-2.6 MeV, finding a W-value of 11.5 eV and a recombination-independent energy scale about 15% below the NEST simulation prediction.