The cepstral method achieves 0.09% strain precision at 1 nm resolution on a small-pixel-count detector, with precision scaling proportional to convergence angle over the square root of dose.
(a) Direct comparison of strain maps calculated from datasets taken with and without precession at the same experimental conditions (dose = 2 pC, α = 2.5 mrad)
1 Pith paper cite this work. Polarity classification is still indexing.
1
Pith paper citing it
fields
physics.ins-det 1years
2025 1verdicts
CONDITIONAL 1representative citing papers
citing papers explorer
-
Cepstral Strain Mapping for Small Pixel-Count Detectors
The cepstral method achieves 0.09% strain precision at 1 nm resolution on a small-pixel-count detector, with precision scaling proportional to convergence angle over the square root of dose.