Retrospective analysis of 302 pediatric CIED surgeries yields weight thresholds (epicardial-only below 21 kg, dual use 21-66 kg, endocardial above) and lead-length constraints to guide realistic MRI heating simulations.
Comparative assessment of RF-induced heating in epicardial implantable electronic devices during 0.55 T and 1.5 T MRI: Effects of full systems vs. abandoned leads,
1 Pith paper cite this work. Polarity classification is still indexing.
1
Pith paper citing it
fields
physics.med-ph 1years
2026 1verdicts
UNVERDICTED 1representative citing papers
citing papers explorer
-
Weight-Guided Constraints for Body Model and Lead Selection in Pediatric CIED MRI Safety Simulations
Retrospective analysis of 302 pediatric CIED surgeries yields weight thresholds (epicardial-only below 21 kg, dual use 21-66 kg, endocardial above) and lead-length constraints to guide realistic MRI heating simulations.