The JUNO calibration house has been installed and tested, maintaining oxygen below 10 ppm and radon below 15 mBq/m3.
Research of radon diffusion behavior in liquid scintillator
1 Pith paper cite this work. Polarity classification is still indexing.
abstract
The background caused by radon and its daughters is an important background in the low background liquid scintillator (LS) detectors. The study of the diffusion behaviour of radon in the LS contributes to the analysis of the related background caused by radon. Methodologies and devices for measuring the diffusion coefficient and solubility of radon in materials are developed and described. The radon diffusion coefficient of the LS was measured for the first time and in addition the solubility coefficient was also obtained. In addition, the radon diffusion coefficient of the polyolefine film which is consistent with data in the literature was measured to verify the reliability of the diffusion device.
citation-role summary
citation-polarity summary
fields
physics.ins-det 1years