pith. machine review for the scientific record. sign in

arxiv: 0809.3665 · v1 · submitted 2008-09-22 · 🌌 astro-ph

Recognition: unknown

Energy-Scale Calibration of the Suzaku X-Ray Imaging Spectrometer Using the Checker Flag Charge Injection Technique in Orbit

Authors on Pith no claims yet
classification 🌌 astro-ph
keywords chargeenergyresolutioncfciinjectionorbitsuzakux-ray
0
0 comments X
read the original abstract

The X-ray Imaging Spectrometer (XIS) on board the Suzaku satellite is an X-ray CCD camera system that has superior performance such as a low background, high quantum efficiency, and good energy resolution in the 0.2-12 keV band. Because of the radiation damage in orbit, however, the charge transfer inefficiency (CTI) has increased, and hence the energy scale and resolution of the XIS has been degraded since the launch of July 2005. The CCD has a charge injection structure, and the CTI of each column and the pulse-height dependence of the CTI are precisely measured by a checker flag charge injection (CFCI) technique. Our precise CTI correction improved the energy resolution from 230 eV to 190 eV at 5.9 keV in December 2006. This paper reports the CTI measurements with the CFCI experiments in orbit. Using the CFCI results, we have implemented the time-dependent energy scale and resolution to the Suzaku calibration database.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.