Mid-infrared VIPA Spectrometer for Rapid and Broadband Trace Gas Detection
read the original abstract
We present and characterize a 2-D imaging spectrometer based on a virtually-imaged phased array (VIPA) disperser for rapid, high-resolution molecular detection using mid-infrared (MIR) frequency combs at 3.1 and 3.8 \mu m. We demonstrate detection of CH4 at 3.1 \mu m with >3750 resolution elements spanning >80 nm with ~600 MHz resolution in a <10 \mu s acquisition time. In addition to broadband detection, rapid, time-resolved single-image detection is demonstrated by capturing dynamic concentration changes of CH4 at a rate of ~375 frames per second. Changes in absorption above the noise floor of 5\times 10-4 are readily detected on the millisecond time scale, leading to important future applications such as real time monitoring of trace gas concentrations and detection of reactive intermediates.
This paper has not been read by Pith yet.
discussion (0)
Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.