pith. machine review for the scientific record. sign in

arxiv: 1206.3146 · v1 · submitted 2012-06-14 · ⚛️ physics.ins-det

Recognition: unknown

Characterisation and calibration of a scintillating fibre detector with > 4000 multi-anode photomultiplier channels

Authors on Pith no claims yet
classification ⚛️ physics.ins-det
keywords detectorautomatedcalibrationchannelsdifferentfibremulti-anodephotomultiplier
0
0 comments X
read the original abstract

In the Kaos spectrometer at the Mainz Microtron a high-resolution coordinate detector for high-energy particles is operated. It consists of scintillating fibres with diameters of < 1mm and is read out by > 4000 multi-anode photomultiplier channels. It is one of the most modern focal-plane detectors for magnetic spectrometers world-wide. To correct variations in the detection efficiency, caused by the different gains and the different optical transmittances, a fully automated off-line calibration procedure has been developed. The process includes the positioning of a radioisotope source alongside the detector plane and the automated acquisition and analysis of the detector signals. It was possible to characterise and calibrate each individual fibre channel with a low degree of human interaction.

This paper has not been read by Pith yet.

discussion (0)