pith. sign in

arxiv: 1308.6018 · v2 · pith:36ZCJKGPnew · submitted 2013-08-27 · 🧬 q-bio.QM

Dynamic Metrics of Heart Rate Variability

classification 🧬 q-bio.QM
keywords metricsdynamicdataresultsacquisitionclassdesigneddevices
0
0 comments X
read the original abstract

Numerous metrics of heart rate variability (HRV) have been described, analyzed, and compared in the literature. However, they rarely cover the actual metrics used in a class of HRV data acquisition devices - those designed primarily to produce real-time metrics. This paper characterizes a class of metrics that we term dynamic metrics. We also report the results of a pilot study which compares one such dynamic metric, based on photoplethysmographic data using a moving sampling window set to the length of an estimated breath cycle (EBC), with established HRV metrics. The results show high correlation coefficients between the dynamic EBC metrics and the established static SDNN metric (standard deviation of Normal-to-Normal) based on electrocardiography. These results demonstrate the usefulness of data acquisition devices designed for real-time metrics.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.