The Highly Miniaturised Radiation Monitor

Authors on Pith no claims yet
classification ⚛️ physics.ins-det
keywords monitorradiationhighlyhmrmminiaturisedachievingactivecalibration
0
0 comments X
read the original abstract

We present the design and preliminary calibration results of a novel highly miniaturised particle radiation monitor (HMRM) for spacecraft use. The HMRM device comprises a telescopic configuration of active pixel sensors enclosed in a titanium shield, with an estimated total mass of 52 g and volume of 15 cm$^3$. The monitor is intended to provide real-time dosimetry and identification of energetic charged particles in fluxes of up to 10$^8$ cm$^{-2}$ s$^{-1}$ (omnidirectional). Achieving this capability with such a small instrument could open new prospects for radiation detection in space.