pith. sign in

arxiv: 1510.02365 · v4 · pith:5NOF75NInew · submitted 2015-10-08 · ⚛️ physics.ins-det

First Measurements with New High-Resolution Gadolinium-GEM Neutron Detectors

classification ⚛️ physics.ins-det
keywords resolutiondetectorsneutronconvertersefficiencyfirstgadoliniuminstruments
0
0 comments X
read the original abstract

European Spallation Source instruments like the macromolecular diffractometer, NMX, require an excellent neutron detection efficiency, high-rate capabilities, time resolution, and an unprecedented spatial resolution in the order of a few hundred micrometers over a wide angular range of the incoming neutrons. For these instruments solid converters in combination with Micro Pattern Gaseous Detectors (MPGDs) are a promising option. A GEM detector with gadolinium converter was tested on a cold neutron beam at the IFE research reactor in Norway. The {\mu}TPC analysis, proven to improve the spatial resolution in the case of $^{10}$B converters, is extended to gadolinium based detectors. For the first time, a Gd-GEM was successfully operated to detect neutrons with a measured efficiency of 11.8% at a wavelength of 2 {\AA} and a position resolution better than 250 {\mu}m.