Calibration of a two-phase xenon time projection chamber with a ³⁷Ar source
classification
⚛️ physics.ins-det
keywords
chargedriftdetectorelectricfieldlightpeaksource
pith:HTSWPVBW Add to your LaTeX paper
What is a Pith Number?\usepackage{pith}
\pithnumber{HTSWPVBW}
Prints a linked pith:HTSWPVBW badge after your title and writes the identifier into PDF metadata. Compiles on arXiv with no extra files. Learn more
read the original abstract
We calibrate a two-phase xenon detector at 0.27 keV in the charge channel and at 2.8 keV in both the light and charge channels using a $^{37}$Ar source that is directly released into the detector. We map the light and charge yields as a function of electric drift field. For the 2.8 keV peak, we calculate the Thomas-Imel box parameter for recombination and determine its dependence on drift field. For the same peak, we achieve an energy resolution, $E_{\sigma}/E_{mean}$, between 9.8% and 10.8% for 0.1 kV/cm to 2 kV/cm electric drift fields.
This paper has not been read by Pith yet.
discussion (0)
Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.