Energy Dependent Features of X-ray Signals in a GridPix Detector
classification
⚛️ physics.ins-det
keywords
energydetectorgridpixresolutionx-rayaboveargonbetter
read the original abstract
We report on the calibration of an argon/isobutane (97.7 %/2.3 %)-filled GridPix detector with soft X-rays (277 eV to 8 keV) using the variable energy X-ray source of the CAST Detector Lab at CERN. We study the linearity and energy resolution of the detector using both the number of hit pixels and the total measured charge as energy measures. For the latter, the energy resolution {\sigma}_E/E is better than 10 % (20 %) for energies above 2 keV (0.5 keV). Several characteristics of the recorded events are studied.