pith. sign in

arxiv: 1802.01319 · v2 · pith:37RGCJV2new · submitted 2018-02-05 · ⚛️ physics.ins-det

Test beam measurement of the first prototype of the fast silicon pixel monolithic detector for the TT-PET project

classification ⚛️ physics.ins-det
keywords tt-petdetectiondetectorfirstmonolithicpixelprototyperesolution
0
0 comments X
read the original abstract

The TT-PET collaboration is developing a PET scanner for small animals with 30 ps time-of-flight resolution and sub-millimetre 3D detection granularity. The sensitive element of the scanner is a monolithic silicon pixel detector based on state-of-the-art SiGe BiCMOS technology. The first ASIC prototype for the TT-PET was produced and tested in the laboratory and with minimum ionizing particles. The electronics exhibit an equivalent noise charge below 600 e- RMS and a pulse rise time of less than 2 ns, in accordance with the simulations. The pixels with a capacitance of 0.8 pF were measured to have a detection efficiency greater than 99% and, although in the absence of the post-processing, a time resolution of approximately 200 ps.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.