pith. sign in

arxiv: 1811.08637 · v1 · pith:LJ35FDARnew · submitted 2018-11-21 · ⚛️ physics.ins-det

High-rate glass MRPC for fixed target experiments at Nuclotron

classification ⚛️ physics.ins-det
keywords mrpccountingexperimentrateacceleratorbeamcapabilitychamber
0
0 comments X
read the original abstract

A Multi-gap Resistive Plate Chamber (MRPC) equipped with heaters to improve the counting rate capability was designed for the BM@N experiment in Dubna. The measurements were performed using a muon beam at IHEP U-70 accelerator in Protvino. The MRPC at 40$^0$C tolerates counting rate up to 6 kHz/cm$^2$ with time resolution ~65 ps and efficiency ~95\% which complies with the conditions of the experiment.