Compact vacuum phototriodes for operation in strong magnetic field
classification
⚛️ physics.ins-det
hep-ex
keywords
fieldmagneticamplitudeoutputphototriodesvacuumanodecompact
read the original abstract
The results of tests of 1'' vacuum phototriodes in a magnetic field up to 4.5 T are presented. It was found that output amplitude decreases by about 6 % per tesla in the magnetic field range from 2.0 to 4.0 T. For devices with an anode mesh pitch of 16$\mu$m, the output amplitude at 4.0 T is 30 % lower than that at zero field.
This paper has not been read by Pith yet.