pith. sign in

arxiv: physics/0106089 · v1 · submitted 2001-06-27 · ⚛️ physics.ins-det

Sub-nanosecond, time-resolved, broadband infrared spectroscopy using synchrotron radiation

classification ⚛️ physics.ins-det
keywords infraredsynchrotronfacilitybroadbandcm-1durationnationalpulses
0
0 comments X
read the original abstract

A facility for sub-nanosecond time-resolved (pump-probe) infrared spectroscopy has been developed at the National Synchrotron Light Source of Brookhaven National Laboratory. A mode-locked Ti:sapphire laser produces 2 ps duration, tunable near-IR pump pulses synchronized to probe pulses from a synchrotron storage ring. The facility is unique on account of the broadband infrared from the synchrotron, which allows the entire spectral range from 2 cm-1 (0.25 meV) to 20,000 cm-1 (2.5 eV) to be probed. A temporal resolution of 200 ps, limited by the infrared synchrotron-pulse duration, is achieved. A maximum time delay of 170 ns is available without gating the infrared detector. To illustrate the performance of the facility, a measurement of electron-hole recombination dynamics for an HgCdTe semiconductor film in the far- and mid infrared range is presented.