Light multi-GEM detector for high-resolution tracking systems
classification
⚛️ physics.ins-det
keywords
detectorfoilsaboveallowsbudgetcontrolledcopperdetectors
read the original abstract
Controlled etching of copper electrodes in Gas Electron Multiplier foils allows a reduction of the material budget by more than a factor of two for a triple-GEM detector. Detectors making use of thinned foils provide performances similar to those obtained with standard devices: a gain above 10^4 for a double-GEM, with energy resolution of 27 % fwhm for 5.9 keV X-rays.
This paper has not been read by Pith yet.