pith. machine review for the scientific record. sign in

arxiv: 1511.03950 · v1 · submitted 2015-11-12 · ⚛️ physics.ins-det

Recognition: unknown

The Practice of Pulse Processing

Authors on Pith no claims yet
classification ⚛️ physics.ins-det
keywords pulsex-rayanalysisdataessentialmicrocalorimeterspracticesteps
0
0 comments X
read the original abstract

The analysis of data from x-ray microcalorimeters requires great care; their excellent intrinsic energy resolution cannot usually be achieved in practice without a statistically near-optimal pulse analysis and corrections for important systematic errors. We describe the essential parts of a pulse-analysis pipeline for data from x-ray microcalorimeters, including steps taken to reduce systematic gain variation and the unwelcome dependence of filtered pulse heights on the exact pulse-arrival time. We find these steps collectively to be essential tools for getting the best results from a microcalorimeter-based x-ray spectrometer.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.

Forward citations

Cited by 1 Pith paper

Reviewed papers in the Pith corpus that reference this work. Sorted by Pith novelty score.

  1. Characterization of a Two-Channel Optical and Near-infrared Transition Edge Sensor System for Rare-Event Searches

    physics.ins-det 2026-05 unverdicted novelty 4.0

    A two-channel TES system for 1064 nm achieves 86% efficiency, <7% energy resolution, and <6 mHz background, allowing 5-sigma detection of signals at 2.7e-5 Hz (5e-24 W) in 20 days.