HiCal 2: An instrument designed for calibration of the ANITA experiment and for Antarctic surface reflectivity measurements

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 1710.11175 v2 pith:IBE754VZ submitted 2017-10-30 physics.ins-det

classification physics.ins-det
keywords calibrationhicalsurfaceanita-4antarcticexperimentflewhical-2
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
read the original abstract

The NASA supported High-Altitude Calibration (HiCal)-2 instrument flew as a companion balloon to the ANITA-4 experiment in December 2016. Based on a HV discharge pulser producing radio-frequency (RF) calibration pulses, HiCal-2 comprised two payloads, which flew for a combined 18 days, covering 1.5 revolutions of the Antarctic continent. ANITA-4 captured over 10,000 pulses from HiCal, both direct and reflected from the surface, at distances varying from 100-800 km, providing a large dataset for surface reflectivity measurements. Herein we present details on the design, construction and performance of HiCal-2.