A modular phantom and software to characterize 3D geometric distortion in MRI

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 2003.10351 v3 pith:NQHSMIZ6 submitted 2020-03-23 physics.med-ph

classification physics.med-ph
keywords phantomdistortionmodularsoftwaregeometricacrosscharacterizecorrection
verification ladder T0 review T1 audit T2 compute T3 formal

Signed reviews

No signed human review yet.

0 comments
read the original abstract

MRI offers outstanding soft tissue contrast that may reduce uncertainties in target and organ-at-risk delineation and enable online adaptive image-guided treatment. Spatial distortions resulting from non-linearities in the gradient fields and non-uniformity in the main magnetic field must be accounted for across the imaging field-of-view to prevent systematic errors during treatment delivery. This work presents a modular phantom and software application to characterize geometric distortion (GD) within the large field-of-view MRI images required for radiation therapy simulation. The modular phantom is assembled from a series of rectangular foam blocks containing high-contrast fiducial markers in a known configuration. The modular phantom design facilitates transportation of the phantom between different MR scanners and MR-guided linear accelerators and allows the phantom to be adapted to fit different sized bores or coils. The phantom was evaluated using a 1.5T MR-guided linear accelerator (MR-Linac) and 1.5T and 3.0T diagnostic scanners. Performance was assessed by varying acquisition parameters to induce image distortions in a known manner. Imaging was performed using T1 and T2 weighted pulse sequences with 2D and 3D distortion correction algorithms and the receiver bandwidth (BW) varied as 250-815 Hz/pixel. Phantom set-up reproducibility was evaluated across independent set-ups. The software was validated by comparison with a non-modular phantom. Average geometric distortion was 0.94+/-0.58 mm for the MR-Linac, 0.90+/-0.53 mm for the 1.5 T scanner, and 1.15+/-0.62 mm for the 3.0T scanner, for a 400 mm diameter volume-of-interest. GD increased, as expected, with decreasing BW, and with the 2D versus 3D correction algorithm. Differences in GD attributed to phantom set-up were 0.13 mm or less. Differences in GD for the two software applications were less than 0.07 mm.

Discussion (0). Continue with ORCID to comment.

Pith tools