REVIEW 2 major objections 4 minor 60 references
Optimisation of Magnetic Field Sensing with Optically Pumped Magnetometers for Magnetic Detection Electrical Impedance Tomography
T0 review · 2 major / 4 minor · reviewed 2026-08-11 · deepseek-v4-flash
Pith's one-line read For MDEIT brain imaging, the best magnetometer configuration is a single axis normal to the scalp, and larger sensors improve image quality.
desk verdict Useful parameter sweep for OPM-based MDEIT, but the cell-volume headline rests on a real arithmetic error in Table 2 and should be re-run. read the letter →
The pith
A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.
The reading
What carries the argument
The argument runs on a simulation pipeline, but the load-bearing physical identity is the fundamental sensitivity limit of an OPM: $\delta B_{\mathrm{opt}} = \frac{\sqrt{2\,BW}}{\gamma \sqrt{nV}}$, which says that for fixed atomic density $n$, bandwidth $BW$, and gyromagnetic ratio $\gamma$, the sensor noise floor falls as the inverse square root of the vapour-cell volume $V$. The paper scales the intrinsic noise for five cell sizes using this formula, adds environmental and current-source noise, embeds the sensors in a finite-element head model, and reconstructs images. Image quality is scored with the weighted spatial variance, a metric that measures how closely the reconstructed conductivity blob matches the true perturbation's location and extent.
What would settle it
Measure the intrinsic noise of OPMs with different vapour-cell volumes while holding atomic density and bandwidth constant; if the noise does not fall as $1/\sqrt{V}$, the predicted image-quality gain from larger cells will not appear. Alternatively, build a phantom MDEIT system with two OPM sizes and check whether reconstructed image quality actually improves with volume as the simulations predict.
Extended reading notes
Core claim
On the paper's own terms, the central discovery is a design rule: for MDEIT, the optimal OPM configuration is a single sensing axis normal to the scalp, and larger vapour cells improve reconstructed image quality. This conclusion comes from finite-element simulations of a seven-tissue human head with four neural perturbations, image reconstruction using 0th-order Tikhonov regularisation with uncorrelated noise-based correction, and assessment using the weighted spatial variance. The authors report that single-axis normal measurements produced the largest signal-to-noise ratio and best image quality, that adding tangential axes introduced noisy measurements that degraded quality, and that increasing vapour-cell side length from 1 mm to 18 mm reduced sensor noise and improved image quality while the rank of the Jacobian stayed constant.
Load-bearing premise
The paper assumes that a larger OPM vapour cell can be operated with the same atomic density and bandwidth, so that its intrinsic noise drops as the inverse square root of the volume; if real larger cells cannot maintain those parameters, the predicted size benefit disappears.
Editorial extensions
If this is right
- Future OPMs built for MDEIT should prioritise a single high-sensitivity axis normal to the scalp over multi-axis capability.
- Arrays of roughly 48 to 96 single-axis sensors are practically suitable; going beyond about 128 gives little additional benefit, and around 32 may suffice for cortical activity.
- Manufacturers can increase vapour-cell volume beyond current commercial sizes to gain sensitivity and bandwidth, potentially approaching SQUID-level performance without cryogenic cooling.
- For a fixed total number of measurements, using more magnetometers is more beneficial than using multi-axis measurements on fewer magnetometers.
Reading between the lines
- If the $1/\sqrt{V}$ noise scaling holds experimentally, the cell-volume benefit could extend to other on-scalp magnetic sensing applications, not just MDEIT, wherever spatial averaging is acceptable.
- The paper shows the Jacobian rank is unchanged by cell size, but real larger cells may also increase standoff distance and power draw; whether those practical penalties offset the sensitivity gain is a question for hardware prototypes.
- One testable extension is to verify the exact noise values in Table 2 against direct measurements of OPMs with different cell volumes, since the table's numbers appear to derive from a single 3 mm reference cell and may not reflect practical operation at constant atomic density and bandwidth.
Signed reviews
Editorial analysis
A structured set of objections, weighed in public.
Referee Report
Summary. This computational modelling study optimises the sensing axis, sensor number, and vapour cell volume of optically pumped magnetometers (OPMs) for Magnetic Detection Electrical Impedance Tomography (MDEIT). Using an anatomically realistic finite-element head model, the authors simulate forward and inverse problems with four neural perturbations at different depths, add realistic noise, and evaluate reconstructed image quality with weighted spatial variance (WSV). They report that single-axis measurements normal to the scalp give the best image quality, that image quality improves with increasing sensor number but with diminishing returns, and that larger OPM vapour cell volumes improve image quality because of the assumed 1/sqrt(V) sensitivity scaling. The paper concludes that future OPMs for MDEIT should be single-axis, highly sensitive, and not necessarily constrained to the sizes of current commercial sensors.
Significance. The paper addresses a practically important hardware-design question for a novel neuroimaging technique, and the computational framework is careful in several respects: the forward model is realistic, the inverse crime is avoided by using different forward and reconstruction meshes, four perturbations at different depths are considered, and statistical testing is performed over 100 noise realisations. The sensing-axis and sensor-number results appear consistent with the described methods and provide useful design guidance. The cell-volume finding, if it survives re-analysis with corrected noise values, would be a significant driver for OPM development. However, the cell-volume result is a direct consequence of the assumed fundamental-sensitivity scaling rather than an independent empirical discovery; the main non-trivial aspect is the demonstration that spatial averaging and standoff effects do not negate the sensitivity gain, as evidenced by the constant Jacobian rank.
major comments (2)
- [Section 2.4, Table 2]
- [Section 4.6 and Eq. (3)]
minor comments (4)
- [References]
- [Throughout text]
- [Section 2.4]
- [Section 4.4]
Circularity Check
No circular derivation: the optimisation results are genuine outputs of a forward/inverse simulation, though the cell-volume conclusion is directly shaped by the assumed 1/sqrt(V) sensitivity scaling.
full rationale
The paper's central claims (single-axis normal best, diminishing returns with sensor count, and image quality rising with OPM cell volume) are outputs of explicit FEM forward/inverse simulations with additive and multiplicative noise models, not fits to the quantities they claim to predict. The cell-volume result is the least independent: Section 2.4 sets the OPM noise from Eq. 3, which has sensitivity proportional to 1/sqrt(V), so the SNR benefit of larger cells is built into the input. However, the paper states this openly, and the simulation still tests the nontrivial question of whether spatial averaging of the forward field over a larger cell destroys the benefit; the rank of the Jacobian and WSV results are not tautological. Self-citations to [5] and [20] are used for the FEM, noise case, and U-NBC reconstruction algorithm; they are disclosed methodological lineage, not load-bearing appeals to an unverified uniqueness theorem or a fitted result. One correctness concern is flagged: Table 2's 'total noise after 232 measurement averages' appears to omit the sqrt(bandwidth) factor used in Section 2.2; for example, 261 fT/rtHz for the 1 mm cell becomes 17.1 fT instead of approximately 542 fT if S times sqrt(1000) divided by sqrt(232) is applied. This would change absolute SNR levels and the balance against current-source noise, although the common multiplicative factor preserves the relative ordering across cell sizes. This is a numerical and validation issue, not a circularity. Overall, no significant circularity is present; the score of 2 reflects the heavy but non-circular reliance on the authors' own prior modelling and reconstruction tools.
Assumptions & free parameters
free parameters (4)
- Bandwidth BW =
500 Hz
- Atomic density n =
1.5e11 mm^-3
- Number of measurement averages =
232
- WSV threshold =
50% of maximum reconstructed conductivity
assumptions (5)
- domain assumption The 7-tissue FEM of the human head is an accurate representation for modelling MDEIT.
- domain assumption The perturbation (1% conductivity increase, 3.86 cm^3) represents fast neural activity.
- standard math OPM sensitivity scales as 1/sqrt(V) with constant n, BW, and gamma (Eq. 3).
- domain assumption The noise model (uncorrelated additive plus correlated multiplicative, Noise Case 1 from [5]) is realistic.
- domain assumption Tikhonov regularization with U-NBC and the WSV metric are appropriate for evaluating image quality.
Cite this review
Pith. "Pith review of Optimisation of Magnetic Field Sensing with Optically Pumped Magnetometers for Magnetic Detection Electrical Impedance Tomography." pith.science (2026). https://pith.science/paper/3WA6N6DC
@misc{pith2026241213354,
author = {Pith},
title = {Pith review of: Optimisation of Magnetic Field Sensing with Optically Pumped Magnetometers for Magnetic Detection Electrical Impedance Tomography},
year = {2026},
howpublished = {\url{https://pith.science/paper/3WA6N6DC}},
note = {Machine review of arXiv:2412.13354}
}
read the original abstract
Magnetic Detection Electrical Impedance Tomography is a novel technique that could enable non-invasive imaging of fast neural activity in the brain. However, commercial magnetometers are not suited to its technical requirements. Computational modelling was used to determine the optimal number, size and orientation of magnetometers, to inform the future development of MDEIT-specific magnetometers. Images were reconstructed using three sensing axes, arrays of 16 to 160 magnetometers, and cell sizes ranging from 1 to 18 mm. Image quality was evaluated visually and with the weighted spatial variance. Single-axis measurements normal to the surface provided the best image quality, and image quality increased with an increase in sensor number and size. This study can inform future OPM design, showing the size of the vapour cell need not be constrained to that of commercially available OPMs, and that a small array of single-axis, highly sensitive sensors is optimal for MDEIT.
Figures
Figures from the paper (2 more)
Reference graph
Works this paper leans on
-
[1]
What we can do and what we cannot do with fMRI,
N. K. Logothetis, “What we can do and what we cannot do with fMRI,” Nature, vol. 453, no. 7197, pp. 869–878, 2008
work page 2008
-
[2]
A. Adler and D. Holder, Electrical Impedance Tomography: Methods, History and Applications , ser. Series in Medical Physics and Biomedical Engineering. CRC Press, 2021. [Online]. Available: https://books.google.co.uk/books?id=5bBZEAAAQBAJ
work page 2021
-
[3]
P. C. Hansen, M. L. Kringelbach, and R. Salmelin, MEG: An Introduction to Methods . Oxford University Press, 2010
work page 2010
-
[4]
K. Mason, K. Aristovich, and D. Holder, “The Feasibility of Fast Neural Magnetic Detection Electrical Impedance Tomography: A Modelling Study,” International IEEE/EMBS Conference on Neural Engineering, NER , vol. 2023-April, pp. 1–4, 2023
work page 2023
-
[5]
——, “Non-invasive imaging of neural activity with magnetic detection electrical impedance tomography (MDEIT): a modelling study,” Physiological Measurement, 2023. [Online]. Available: https://iopscience.iop.org/ article/10.1088/2053-1583/abe778
-
[6]
R. H. Ireland, J. C. Tozer, A. T. Barker, and D. C. Barber, “Towards magnetic detection electrical impedance tomography: Data acquisition and image reconstruction of current density in phantoms and in vivo,”Physiological Measurement, vol. 25, no. 3, pp. 775–796, 2004
work page 2004
-
[7]
M. Faulkner, S. Hannan, K. Aristovich, J. Avery, and D. Holder, “Feasibility of imaging evoked activity throughout the rat brain using electrical impedance tomography,” NeuroImage, vol. 178, no. February, pp. 1–10,
-
[8]
O. Gilad and D. S. Holder, “Impedance changes recorded with scalp electrodes during visual evoked responses: Implications for Electrical Impedance Tomography of fast neural activity,” pp. 514–522, 2009
work page 2009
Show all 60 references
-
[9]
Characterising the frequency response of impedance changes during evoked physiological activity in the rat brain,
M. Faulkner, S. Hannan, K. Aristovich, J. Avery, and D. Holder, “Characterising the frequency response of impedance changes during evoked physiological activity in the rat brain,”Physiological Measurement, vol. 39, no. 3, p. 034007, 4 2018. [Online]. Available: https://doi.org...
2018 doi
-
[10]
Elekta Neuromag®,
E. Neuromag and E. Neuromag, “Elekta Neuromag®,” 2008
2008
-
[11]
Ultra-sensitive SQUID systems for applications in biomagnetism and ultra-low field MRI,
R. Körber, O. Kieler, P. Hömmen, N. Höfner, and J. H. Storm, “Ultra-sensitive SQUID systems for applications in biomagnetism and ultra-low field MRI,” in ISEC 2019 - International Superconductive Electronics Conference , 2019, pp. 18–20
2019
-
[12]
DC SQUID Series Array Amplifiers With 120 MHZ Bandwidth (Corrected),
M. E. Huber, P. A. Neil, R. G. Benson, D. A. Bums, A. Corey, C. S. Flynn, Y . Kitaygoradskaya, O. Massihzaueh, J. M. Martinis, and G. Chilton, “DC SQUID Series Array Amplifiers With 120 MHZ Bandwidth (Corrected),” IEEE Transactions on Applied Superconductivity, vol. 11, no. 2,...
2001
-
[13]
Ultra-low-noise EEG/MEG systems enable bimodal non-invasive detection of spike-like human somatosensory evoked responses at 1 kHz,
T. Fedele, H. J. Scheer, M. Burghoff, G. Curio, and R. Körber, “Ultra-low-noise EEG/MEG systems enable bimodal non-invasive detection of spike-like human somatosensory evoked responses at 1 kHz,”Physiological Measurement, vol. 36, no. 2, pp. 357–368, 2015
2015
-
[14]
An ultra-sensitive and wideband magnetometer based on a superconducting quantum interference device,
J. H. Storm, P. Hömmen, D. Drung, and R. Körber, “An ultra-sensitive and wideband magnetometer based on a superconducting quantum interference device,” Applied Physics Letters, vol. 110, no. 7, 2017. [Online]. Available: http://dx.doi.org/10.1063/1.4976823
2017 doi
-
[15]
A modular, extendible and field-tolerant multichannel vector magnetometer based on current sensor squids,
J. H. Storm, D. Drung, M. Burghoff, and R. Körber, “A modular, extendible and field-tolerant multichannel vector magnetometer based on current sensor squids,” Superconductor Science and Technology, vol. 29, 7 2016
2016
-
[16]
Squid current sensors with an integrated thermally actuated input current limiter,
R. Körber, P. Krzysteczko, M. Klemm, T. Liu, J. H. Storm, and J. Beyer, “Squid current sensors with an integrated thermally actuated input current limiter,” Superconductor Science and Technology, vol. 36, 7 2023
2023
-
[17]
Magnetoen- cephalogram systems developed at kit,
H. Kado, M. Higuchi, M. Shimogawara, Y . Haruta, Y . Adachi, J. Kawai, H. Ogata, and G. Uehara, “Magnetoen- cephalogram systems developed at kit,” 1999
1999
-
[18]
The atomic magnetometer: A new era in biomagnetism,
R. T. Wakai, “The atomic magnetometer: A new era in biomagnetism,”AIP Conference Proceedings, vol. 1626, pp. 46–54, 2014
2014
-
[19]
Magnetocardiography measurements with 4He vector optically pumped magnetometers at room temperature,
S. Morales, M. C. Corsi, W. Fourcault, F. Bertrand, G. Cauffet, C. Gobbo, F. Alcouffe, F. Lenouvel, M. Le Prado, F. Berger, G. Vanzetto, and E. Labyt, “Magnetocardiography measurements with 4He vector optically pumped magnetometers at room temperature,” Physics in Medicine and...
2017
-
[20]
Noise-based correction for electrical impedance tomography,
K. Mason, F. Maurino-Alperovich, D. Holder, and K. Aristovich, “Noise-based correction for electrical impedance tomography,” Physiological Measurement, vol. 45, no. 6, p. 065002, 2024. 14 Optimisation of OPMs for MDEIT
2024
-
[21]
Simplifying the hardware requirements for fast neural EIT of peripheral nerves,
E. Ravagli, S. Mastitskaya, D. Holder, and K. Aristovich, “Simplifying the hardware requirements for fast neural EIT of peripheral nerves,” Physiological Measurement, vol. 43, no. 1, 2022
2022
-
[22]
Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalography,
T. M. Tierney, N. Holmes, S. Mellor, J. D. López, G. Roberts, R. M. Hill, E. Boto, J. Leggett, V . Shah, M. J. Brookes, R. Bowtell, and G. R. Barnes, “Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalography,” NeuroImage, vol. 199, no. Decem...
2018
-
[23]
P. C. Hansen, Discrete inverse problems: insight and algorithms . SIAM, 2010
2010
-
[24]
Optimising the sensing volume of OPM sensors for MEG source reconstruction,
Y . Bezsudnova, L. M. Koponen, G. Barontini, and O. Jensen, “Optimising the sensing volume of OPM sensors for MEG source reconstruction,” NeuroImage, vol. 264, no. June, pp. 1–11, 2022. [Online]. Available: https://doi.org/10.1016/j.neuroimage.2022.119747
2022
-
[25]
QZFM, QuSpin,
Quspin, “QZFM, QuSpin,” 2022. [Online]. Available: https://quspin.com/products-qzfm/
2022
-
[26]
Fieldline,
F. Inc., “Fieldline,” 2023. [Online]. Available: https://fieldlineinc.com/
2023
-
[27]
Mag4health,
MAG4Health, “Mag4health,” https://www.mag4health.com/product/, 2021
2021
-
[28]
Multi-channel whole-head OPM-MEG: Helmet design and a comparison with a conventional system,
R. M. Hill, E. Boto, M. Rea, N. Holmes, J. Leggett, L. A. Coles, M. Papastavrou, S. K. Everton, B. A. Hunt, D. Sims, J. Osborne, V . Shah, R. Bowtell, and M. J. Brookes, “Multi-channel whole-head OPM-MEG: Helmet design and a comparison with a conventional system,” NeuroImage, ...
2020
-
[29]
A 20-channel magnetoencephalography system based on optically pumped magnetometers,
A. Borna, T. R. Carter, J. D. Goldberg, A. P. Colombo, Y . Y . Jau, C. Berry, J. McKay, J. Stephen, M. Weisend, and P. D. Schwindt, “A 20-channel magnetoencephalography system based on optically pumped magnetometers,” Physics in Medicine and Biology , vol. 62, no. 23, pp. 8909...
2017
-
[30]
How to build a magnetometer with thermal atomic vapor: a tutorial,
A. Fabricant, I. Novikova, and G. Bison, “How to build a magnetometer with thermal atomic vapor: a tutorial,” New Journal of Physics, vol. 25, no. 2, p. 025001, 2023
2023
-
[31]
Magnetoencephalography with optically pumped magnetometers (OPM-MEG): the next generation of functional neuroimaging,
M. J. Brookes, J. Leggett, M. Rea, R. M. Hill, N. Holmes, E. Boto, and R. Bowtell, “Magnetoencephalography with optically pumped magnetometers (OPM-MEG): the next generation of functional neuroimaging,” Trends in Neurosciences, vol. 45, no. 8, pp. 621–634, 2022. [Online]. Avai...
2022 doi
-
[32]
Four-channel optically pumped atomic magnetometer for magnetoencephalography,
A. P. Colombo, T. R. Carter, A. Borna, Y .-Y . Jau, C. N. Johnson, A. L. Dagel, and P. D. D. Schwindt, “Four-channel optically pumped atomic magnetometer for magnetoencephalography,” Optics Express, vol. 24, no. 14, p. 15403, 2016
2016
-
[33]
An integrated full-head OPM-MEG system based on 128 zero-field sensors,
O. Alem, K. J. Hughes, I. Buard, T. P. Cheung, T. Maydew, A. Griesshammer, K. Holloway, A. Park, V . Lechuga, C. Coolidge, M. Gerginov, E. Quigg, A. Seames, E. Kronberg, P. Teale, and S. Knappe, “An integrated full-head OPM-MEG system based on 128 zero-field sensors,” Frontier...
2023
-
[34]
Grosz, M
A. Grosz, M. J. Haji-Sheikh, and S. C. Mukhopadhyay, High Sensitivity Magnetometers. Springer, 2017, vol. 19. [Online]. Available: http://www.springer.com/series/10617
2017
-
[35]
Magnetoencephalography with optically pumped 4 He magnetometers at ambient temperature,
E. Labyt, M. C. Corsi, W. Fourcault, A. Palacios Laloy, F. Bertrand, F. Lenouvel, G. Cauffet, M. Le Prado, F. Berger, and S. Morales, “Magnetoencephalography with optically pumped 4 He magnetometers at ambient temperature,” IEEE Transactions on Medical Imaging, vol. 38, no. 1,...
2019
-
[36]
Budker and D
D. Budker and D. F. J. Kimball, Optical Magnetometry. Cambridge University Press, 2004, vol. 39, no. 2
2004
-
[37]
Can a quantum nondemolition measurement improve the sensitivity of an atomic magnetometer?
M. Auzinsh, D. Budker, D. F. Kimball, S. M. Rochester, J. E. Stalnaker, A. O. Sushkov, and V . V . Yashchuk, “Can a quantum nondemolition measurement improve the sensitivity of an atomic magnetometer?” Physical Review Letters, vol. 93, no. 17, pp. 1–4, 2004
2004
-
[38]
High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation,
J. C. Allred, R. N. Lyman, T. W. Kornack, and M. V . Romalis, “High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation,” Physical Review Letters, vol. 89, no. 13, pp. 1 308 011–1 308 014, 2002
2002
-
[39]
Towards a miniature atomic scalar magnetometer using a liquid crystal polarization rotator,
J. Rutkowski, W. Fourcault, F. Bertrand, U. Rossini, S. Gétin, S. Le Calvez, T. Jager, E. Herth, C. Gorecki, M. Le Prado, J. M. Léger, and S. Morales, “Towards a miniature atomic scalar magnetometer using a liquid crystal polarization rotator,” Sensors and Actuators, A: Physic...
2014 doi
-
[40]
A compact, high performance atomic magnetometer for biomedical applications,
V . K. Shah and R. T. Wakai, “A compact, high performance atomic magnetometer for biomedical applications,” Physics in Medicine and Biology , vol. 58, no. 22, pp. 8153–8161, 2013
2013
-
[41]
Fully integrated standalone zero field optically pumped magnetometer for biomagnetism,
J. Osborne, J. Orton, O. Alem, and V . Shah, “Fully integrated standalone zero field optically pumped magnetometer for biomagnetism,” in Steep dispersion engineering and opto-atomic precision metrology XI , vol. 10548. SPIE, 2018, pp. 89–95. 15 Optimisation of OPMs for MDEIT
2018
-
[42]
A 90-channel triaxial magnetoencephalography system using optically pumped magnetometers,
M. Rea, E. Boto, N. Holmes, R. Hill, J. Osborne, N. Rhodes, J. Leggett, L. Rier, R. Bowtell, V . Shah, and M. J. Brookes, “A 90-channel triaxial magnetoencephalography system using optically pumped magnetometers,”Annals of the New York Academy of Sciences , vol. 1517, no. 1, p...
2022
-
[43]
Helium-4 magnetometers for room-temperature biomedical imaging: toward collective operation and photon-noise limited sensitivity,
W. Fourcault, R. Romain, G. Le Gal, F. Bertrand, V . Josselin, M. Le Prado, E. Labyt, and A. Palacios-Laloy, “Helium-4 magnetometers for room-temperature biomedical imaging: toward collective operation and photon-noise limited sensitivity,” Optics Express, vol. 29, no. 10, p. ...
2021
-
[44]
Characterization of noise sources in a microfabricated single-beam zero-field optically-pumped magnetometer,
S. P. Krzyzewski, A. R. Perry, V . Gerginov, and S. Knappe, “Characterization of noise sources in a microfabricated single-beam zero-field optically-pumped magnetometer,” Journal of Applied Physics, vol. 126, no. 4, 2019
2019
-
[45]
Subpicotesla atomic magnetometry with a microfabricated vapour cell,
V . Shah, S. Knappe, P. D. Schwindt, and J. Kitching, “Subpicotesla atomic magnetometry with a microfabricated vapour cell,” Nature Photonics, vol. 1, no. 11, pp. 649–652, 2007
2007
-
[46]
Preliminary studies in imaging neuronal depolarization in the brain with Electrical or Magnetic Detection Impedance Tomography,
O. Gilad, “Preliminary studies in imaging neuronal depolarization in the brain with Electrical or Magnetic Detection Impedance Tomography,” Ph.D. dissertation, University College London, 2007
2007
-
[47]
Human Cerebral Activation during Steady-State Visual-Evoked Responses,
M. A. Pastor, J. Artieda, J. Arbizu, M. Valencia, and J. C. Masdeu, “Human Cerebral Activation during Steady-State Visual-Evoked Responses,”Journal of Neuroscience, vol. 23, no. 37, pp. 11 621–11 627, 2003
2003
-
[48]
Nowinski, Introduction to Brain Anatomy
W. Nowinski, Introduction to Brain Anatomy. Springer, 07 2011
2011
-
[49]
Liston, R
A. Liston, R. Bayford, and D. Holder, “A cable theory based biophysical model of resistance change in crab peripheral nerve and human cerebral cortex during neuronal depolarisation: Implications for electrical impedance tomography of fast neural activity in the brain,” pp. 425...
2012 doi
-
[50]
Optimisation of current injection protocol based on a region of interest,
M. Faulkner, M. Jehl, K. Aristovich, J. Avery, A. Witkowska-Wrobel, and D. Holder, “Optimisation of current injection protocol based on a region of interest,” Physiological Measurement, vol. 38, no. 6, pp. 1158–1175, 2017
2017
-
[51]
EIT reconstruction algorithms: Pitfalls, challenges and recent developments,
W. R. Lionheart, “EIT reconstruction algorithms: Pitfalls, challenges and recent developments,”Physiological Measurement, vol. 25, no. 1, pp. 125–142, 2004
2004
-
[52]
Imaging fast electrical activity in the brain with electrical impedance tomography,
K. Y . Aristovich, B. C. Packham, H. Koo, G. S. dos Santos, A. McEvoy, and D. S. Holder, “Imaging fast electrical activity in the brain with electrical impedance tomography,” NeuroImage, vol. 124, pp. 204–213, 1 2016. [Online]. Available: url{http://creativecommons.org/license...
2016
-
[53]
Are patient specific meshes required for eit head imaging ?
M. Jehl, K. Aristovich, and M. Faulkner, “Are patient specific meshes required for eit head imaging ?” Physiologi- cal Measurement, 2016
2016
-
[54]
A versatile and reproducible multi-frequency electrical impedance tomography system,
J. Avery, T. Dowrick, M. Faulkner, N. Goren, and D. Holder, “A versatile and reproducible multi-frequency electrical impedance tomography system,” Sensors (Switzerland), vol. 17, no. 2, 2017
2017
-
[55]
Comsol multiphysics,
COMSOL AB, “Comsol multiphysics,” 2022. [Online]. Available: www.comsol.com
2022
-
[56]
Report of the committee on methods of clinical examination in electroencephalography. 1957,
H. Jasper, “Report of the committee on methods of clinical examination in electroencephalography. 1957,” in Electroencephalography and Clinical Neurophysiology, vol. 10. Elsevier, 5 1958, pp. 370–375
1957
-
[57]
Some Novel Approaches in Modelling and Image Reconstruction for Multi-Frequency Electrical Impedance Tomography of the Human Brain,
L. Horesh, “Some Novel Approaches in Modelling and Image Reconstruction for Multi-Frequency Electrical Impedance Tomography of the Human Brain,” Ph.D. dissertation, University College London, 2006
2006
-
[58]
Spin-exchange relaxation-free magnetometry using elliptically polarized light,
V . Shah and M. V . Romalis, “Spin-exchange relaxation-free magnetometry using elliptically polarized light,” Physical Review A - Atomic, Molecular , and Optical Physics, vol. 80, 8 2009
2009
-
[59]
An exhaustive criterion for estimating quality of images in electrical impedance tomography with application to clinical imaging,
A. Javaherian, A. Movafeghi, R. Faghihi, and E. Yahaghi, “An exhaustive criterion for estimating quality of images in electrical impedance tomography with application to clinical imaging,” Journal of Visual Communication and Image Representation , vol. 24, no. 7, pp. 773–785, ...
2013 doi
-
[2018]
Available: https://doi.org/10.1016/j.neuroimage.2018.05.022
[Online]. Available: https://doi.org/10.1016/j.neuroimage.2018.05.022
2018 doi
Reviewed August 11, 2026 · model on record in the stance chip above.
Discussion (0). Continue with ORCID to comment.