REVIEW 4 major objections 4 minor 51 references
First experimental test of a coded-mask gamma camera for proton therapy monitoring
T0 review · 4 major / 4 minor · reviewed 2026-08-10 · deepseek-v4-flash
Pith's one-line read A coded-mask gamma camera, tested for the first time under clinical proton-therapy conditions, can locate the distal falloff of the prompt-gamma depth profile — and hence the proton range — with a statistical precision of about 1.7 mm for…
desk verdict The first full-scale coded-mask gamma camera test at a clinical proton facility is a real step forward, but the headline 1.7 mm precision at 10^8 protons is not actually derived in the text. read the letter →
The pith
A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.
The reading
What carries the argument
The central object is the coded-mask camera: a structured tungsten collimator with a modified uniformly redundant array (MURA) pattern that casts a position-dependent shadow of the prompt-gamma source onto a pixelated scintillating-fiber detector. The reconstruction machinery is maximum-likelihood expectation maximization (MLEM), iteratively inverting the recorded hit map against a simulated system matrix to recover the one-dimensional prompt-gamma depth profile, from which the distal falloff position is read at half maximum of the falling edge. The system matrix is the load-bearing link between detector hits and source depth: it encodes, for each depth bin, the probability that a prompt gamma is registered in each detector pixel, corrected per-pixel for measured detection efficiency.
What would settle it
Take the same camera into the same geometry but irradiate the PMMA phantom at an energy not used in building the system matrix (say 75 MeV), reconstruct the distal falloff position, and compare with the PSTAR range; if the residual grows beyond the claimed 1.7 mm or the slope departs from the tested energies' fit, the simulated system matrix is not transferable.
Extended reading notes
Core claim
In this experiment, a prototype coded-mask camera — a 7-layer stack of LYSO:Ce,Ca scintillating fibers read out by silicon photomultipliers, shadowed by a 476-rank MURA tungsten mask — was placed beside a PMMA phantom irradiated by proton beams of seven energies from 70.51 to 108.15 MeV at the Heidelberg Ion Therapy Center. Prompt-gamma depth profiles were reconstructed with an MLEM algorithm whose system matrix came from Geant4 simulations of point-source responses. The distal falloff position extracted from each profile tracked the PSTAR-calculated proton range with a Pearson correlation of 0.996 and an RMSE of 1.7 mm; for the reference spot S4 at 90.86 MeV and $10^{8}$ protons, the statistical precision of the distal falloff determination was 1.7 mm (1σ). The agreement with simulations, which reproduce the experimental profiles including artifacts, supports the conclusion that the camera works as modeled. The paper interprets this as validating coded-mask imaging as a competitive alternative for online range verification.
Load-bearing premise
The reconstructed depth profile is only as trustworthy as the simulated system matrix: Monte Carlo point-source responses with an energy spectrum from one beam energy are assumed to hold for all seven beam energies, all lateral positions, and the real detector's efficiency pattern, so any depth- or energy-dependent simulation bias becomes a range-measurement bias.
Editorial extensions
If this is right
- A coded-mask camera can determine proton range shifts in a phantom with about 2 mm precision per 10^8-proton spot, putting it on par with existing slit-based prompt-gamma cameras.
- The detector's rate capability (up to 3.2×10^9 protons/s without dead time, with headroom for higher rates) is sufficient for synchrotron and cyclotron clinical beam intensities, so the approach is not limited by count-rate saturation.
- Simulations of a fully operational detector with no dead pixels predict a fourfold improvement in the RMSE of distal falloff determination (1.7 mm to 0.4 mm), identifying hardware repair as the direct path to sub-millimetre precision.
- The camera is insensitive to lateral beam position changes within ±1 cm at fixed depth, so small patient setup shifts do not corrupt range readings.
- Once y-position resolution is restored, the same detector geometry can be extended to 2D prompt-gamma imaging, as shown in prior simulations.
Reading between the lines
- The 1.7 mm precision is established in a homogeneous PMMA phantom; clinical tissue has density and composition variations that will add systematic errors, so the clinically achievable precision may be worse.
- The system matrix is built from a single beam energy's gamma spectrum and assumed valid across energies; measuring the camera's response at untested energies or with a radioactive line source would directly test this transferability.
- If the simulated factor-of-four improvement is realized after fixing dead pixels, coded-mask cameras could reach sub-millimetre range precision per spot, which would make margin reduction a quantitative trade-off rather than a safety risk.
- The paper's comparison suggests coded masks offer a larger field of view and smaller material budget than slit cameras at similar precision; a head-to-head clinical study would be needed to see whether this translates into better patient outcomes.
Signed reviews
Editorial analysis
A structured set of objections, weighed in public.
Referee Report
Summary. The manuscript reports the first experimental test of a coded-mask (CM) gamma camera for proton therapy monitoring. The system uses a scintillating-fiber detector with a tungsten MURA mask, tested at the Heidelberg Ion Therapy Center with a PMMA phantom and proton beams of seven energies (70.51–108.15 MeV). Prompt-gamma depth profiles are reconstructed with MLEM, and the distal falloff position (DFP) is compared with PSTAR proton ranges. The authors report a statistical precision of DFP determination of 1.7 mm for 10^8 protons at 90.86 MeV, rate capability at clinical beam intensities, and benchmarking against Geant4 simulations. They conclude that the CM camera is a competitive alternative to slit-based systems.
Significance. If the quantitative claim is correct, this would be the first experimental demonstration that a coded-mask camera can track proton range shifts with precision comparable to knife-edge slit and multi-parallel slit systems in a clinically realistic setting. The paper has notable strengths: a seven-energy experimental dataset, rate tests up to 3.2×10^9 protons/s, transparent handling of dead pixels and background, and a detailed simulation chain with optical-photon tracking. However, the main precision claim is not consistently derived in the text: the only 1.7 mm value computed is an RMSE accuracy metric for full-statistics data, not a per-spot statistical precision from the bootstrap analysis. In addition, Table 3 includes a precision value for prior simulation work that the authors themselves retract in footnote 5. The qualitative feasibility result is credible, but the quantitative headline needs substantial revision or re-derivation.
major comments (4)
- [Abstract; Section 3.4; Section 3.5; Section 5] The paper claims a statistical precision of 1.7 mm for 10^8 protons at the reference energy of 90.86 MeV, but the only 1.7 mm value derived in the text is the RMSE computed in Section 3.4 (Eq. 2) from full 10^10-proton profiles across all seven beam energies. The bootstrap analysis in Section 2.5.5 and Figures 10 and 13 is never summarized as a standard deviation for spot S4 at 10^8 protons; Section 3.5 reports only interquartile ranges. The headline precision therefore appears to be a mislabeled accuracy metric rather than a statistically derived per-spot precision. Because this number is the basis of the abstract, the conclusions, and the comparison in Table 3, the authors must either provide the actual bootstrap standard deviation for S4 at 10^8 protons or reword the claims to refer to RMSE accuracy; footnote 5 shows that a similar conflation previously led to an incorrect published value.
- [Section 2.5.4; Section 3.4] The reconstruction parameters (MLEM iteration count, energy threshold, Gaussian smoothing kernel, excluded detector columns, and event classes) are optimized using the same experimental data set on which the RMSE, correlation, and slope are then reported. No held-out validation is described. Because the optimization explicitly targets low RMSE against PSTAR ranges, the quoted RMSE of 1.7 mm is an in-sample fit statistic and may be optimistic. I recommend a cross-validation scheme, such as leave-one-spot-out, or at least an explicit description of how the optimization was prevented from overfitting, to demonstrate that the reported performance is not an artifact of parameter tuning.
- [Table 3; Section 4.2, footnote 5] Table 3 lists a precision of 0.72 mm for CM simulation [25], but footnote 5 in Section 4.2 explicitly retracts the 0.7–1.3 mm range from that prior work as incorrect, stating that it corresponded to aggregated range-shift precision rather than single-spot statistical precision. Using the retracted value in the comparison table is misleading and undermines the claim that the current experimental result is consistent with prior simulation predictions. The entry should be removed, corrected, or annotated with the retraction.
- [Section 2.3.3; Section 3.6] The system matrix A used in the MLEM update (Eq. 1) is generated entirely from Monte Carlo simulations of point-like gamma sources, using the S4 energy spectrum, and the same simulation chain is then validated by its agreement with the experimental DFP data (Table 2). Consequently, the agreement between reconstructed and PSTAR ranges could reflect internal consistency of the simulation rather than the camera's physical response. I ask the authors to provide a sensitivity analysis with respect to the assumed gamma energy spectrum and the detector efficiency model, or to cross-validate using a system matrix computed from an independent simulation or analytical model, to establish that the DFP reconstruction is robust to simulation uncertainties.
minor comments (4)
- [Abstract] The abstract contains typographical and formatting issues, such as "to108.15 MeV" (missing space) and "10 8" instead of "10^8" in several places; please correct these throughout.
- [Section 3.5] The sentence "For all but the deepest beam spot (S1-S6), the IQR is below 3 mm" is confusing because S1-S6 is not a single spot; presumably it means all spots except S7, which is the deepest. Please reword for clarity.
- [Section 4.2, first paragraph; Figure 13] The text states that the statistical-precision studies are "translated to the usual metric of standard deviation (1σ)" and summarized in Fig. 13, but Figure 13 shows no numeric standard deviations. A table of the 1σ values, especially for spot S4 at 10^8 protons, would be necessary to support the claimed 1.7 mm precision.
- [Section 3.1; Section 4.3] The y-position resolution of 74 ± 10 mm FWHM is very poor and the authors note this precludes 2D imaging. The abstract and conclusions should perhaps qualify that the current prototype provides 1D imaging only, to avoid overstating the system's current capabilities.
Circularity Check
No definitional circularity: the central DFP result is an empirical measurement benchmarked against external PSTAR ranges; only minor self-consistency and metric-traceability caveats remain.
full rationale
The paper's core claim is an experimental measurement: DFP positions are extracted from measured hit maps via MLEM and compared with PSTAR proton ranges. PSTAR is an external benchmark, so the 1.7 mm accuracy/precision result does not reduce by construction to the simulation inputs. The MLEM system matrix A is generated by Geant4 simulations and includes an efficiency correction epsilon_i = (R_i - B_i)/SR_i that uses simulated reference data, creating a calibration loop, but the PSTAR comparison breaks any definitional circularity. The bootstrap analysis measures an internal spread of reconstructed DFPs rather than a fitted parameter renamed as a prediction. Two caveats keep the score at 2 rather than 0: the 1.7 mm value appears in Section 3.4 as the 10^10-proton RMSE and in the abstract/conclusions as the 10^8-proton single-spot precision, while Section 3.5 reports only IQRs and Fig. 13, so the text never states the S4/10^8 bootstrap standard deviation that would directly license the headline; and footnote 5 retracts a prior self-cited precision range that still appears in Table 3 as the CM-simulation comparator. These are reporting and self-citation inconsistencies, not equation-level reductions, so no circular step is identified.
Assumptions & free parameters
free parameters (5)
- MLEM iteration count =
23
- Lower hit-map energy threshold =
1 MeV
- Gaussian smoothing kernel standard deviation =
3 pixels
- Excluded lateral fiber columns =
columns 0-2 and 52-54 of 55
- Included event classes =
types 1-4 (unique and semi-unique), type 5 excluded
assumptions (6)
- domain assumption Geant4 QGSP_BIC_HP_EMZ plus GODDeSS optical simulation accurately predicts prompt-gamma production and detector response for 70-108 MeV protons in PMMA.
- domain assumption The system matrix built from point sources using the S4 gamma energy spectrum is valid for all seven beam energies and all lateral beam positions.
- domain assumption The distal falloff position defined as the half-maximum point of a spline fit on the falling edge is a faithful proxy for proton range.
- domain assumption PSTAR-calculated proton ranges in PMMA are accurate ground truth for the tested energies.
- domain assumption The efficiency correction using the 68Ge reference measurement and background subtraction removes all relevant detector nonuniformities and activation background.
- domain assumption Background measured at least 30 minutes after the last beam run represents the inter-spill background during irradiation.
Cite this review
Pith. "Pith review of First experimental test of a coded-mask gamma camera for proton therapy monitoring." pith.science (2026). https://pith.science/paper/7GJYZJNE
@misc{pith2026250100666,
author = {Pith},
title = {Pith review of: First experimental test of a coded-mask gamma camera for proton therapy monitoring},
year = {2026},
howpublished = {\url{https://pith.science/paper/7GJYZJNE}},
note = {Machine review of arXiv:2501.00666}
}
abstract
Objective. The objective of the presented study was to evaluate the feasibility of a coded-mask (CM) gamma camera for real-time range verification in proton therapy, addressing the need for a precise and efficient method of treatment monitoring. Approach. A CM gamma camera prototype was tested in clinical conditions. The setup incorporated a scintillator-based detection system and a structured tungsten collimator. The experiment consisted of the irradiation of PMMA phantom with proton beams of energies ranging from 70.51 to 108.15 MeV. Experimental data were benchmarked against Monte Carlo simulations. The distal falloff position was determined for both experimental data and simulations. Main results. The tested CM camera achieved a statistical precision of distal falloff position determination of 1.7 mm for $10^8$ protons, which is consistent with simulation predictions, despite hardware limitations such as non-functional detector pixels. Simulations indicated that a fully operational setup would further improve the performance of the detector. The system demonstrated rate capability sufficient for clinical proton beam intensities and maintained performance without significant dead time. Significance. This study validates the potential of the CM gamma camera for real-time proton therapy monitoring. The technology promises to enhance treatment accuracy and patient safety, offering a competitive alternative to existing approaches such as single-slit and multi-slit systems.
Figures
Figures from the paper (10 more)
Reference graph
Works this paper leans on
-
[25]
Near-field coded-mask technique and its potential for proton therapy monitoring
R. Hetzel et al. “Near-field coded-mask technique and its potential for proton therapy monitoring”.Physics in Medicine and Biology, 68, 245028 (2023). doi:10.1088/1361-6560/ad05b2
-
[1]
https://www.ptcog.site/, Ac- cessed: 2024-10-21
Proton Therapy Co-Operative Group. https://www.ptcog.site/, Ac- cessed: 2024-10-21
work page 2024
-
[2]
NuPECC report 2014: Nuclear Physics for Medicine
“NuPECC report 2014: Nuclear Physics for Medicine”. http://www. nupecc.org/pub/npmed2014.pdf. Accessed: 2021-07-13
work page 2014
-
[3]
Prompt gamma measurements for locating the dose falloff region in the proton therapy
C.-H. Minet al.“Prompt gamma measurements for locating the dose falloff region in the proton therapy”. Applied Physics Letters, 89(18) (2006). doi:10.1063/1.2378561
-
[4]
Proton range verification through prompt gamma-ray spectroscopy
J. M. Verburg and J. Seco. “Proton range verification through prompt gamma-ray spectroscopy”.Physics in Medicine and Biology, 59(23), 7089 (2014). doi:10.1088/0031-9155/59/23/7089
-
[5]
Absolute prompt-gamma yield measurements for ion beam therapy monitoring
M. Pintoet al.“Absolute prompt-gamma yield measurements for ion beam therapy monitoring”.Physics in Medicine and Biology, 60(2), 565 (2015). doi:10.1088/0031-9155/60/2/565
-
[6]
Spectroscopic study of prompt-gamma emission for range verification in proton therapy
L. Kelleter et al. “Spectroscopic study of prompt-gamma emission for range verification in proton therapy”. Physica Medica, 34, 7 (2017). doi:10.1016/j.ejmp.2017.01.003
-
[7]
Prompt-gamma imaging in particle therapy
M. Pinto. “Prompt-gamma imaging in particle therapy” (2024). doi:10.1140/epjp/s13360-024-05664-4. 24
Show all 51 references
-
[8]
First clinical application of a prompt gamma based in vivo proton range verification system
C. Richteret al.“First clinical application of a prompt gamma based in vivo proton range verification system”.Radiotherapy and Oncology,118(2), 232 (2016). doi:10.1016/j.radonc.2016.01.004
2016 doi
-
[9]
First-in-human validation of CT-based proton range prediction using prompt gamma imaging in prostate cancer treatments
J. Berthold et al. “First-in-human validation of CT-based proton range prediction using prompt gamma imaging in prostate cancer treatments”. International Journal of Radiation Oncology Biology Physics, 111, 1033 (2021). doi:10.1016/j.ijrobp.2021.06.036
2021 doi
-
[10]
Potential margin reduction in prostate cancer pro- ton therapy with prompt gamma imaging for online treatment ver- ification
S. Bertschi et al. “Potential margin reduction in prostate cancer pro- ton therapy with prompt gamma imaging for online treatment ver- ification”. Physics and Imaging in Radiation Oncology , 26 (2023). doi:10.1016/j.phro.2023.100447
2023
-
[11]
Prompt gamma imaging for in vivo range verification of pencil beam scanning proton therapy
Y. Xieet al.“Prompt gamma imaging for in vivo range verification of pencil beam scanning proton therapy”.International Journal of Radiation Oncol- ogy, Biology, Physics, 99(1) (2017). doi:10.1016/j.ijrobp.2017.04.027
2017 doi
-
[13]
Multi-slit prompt-gamma camera for locating of distal dose falloff in proton therapy
J. H. Parket al. “Multi-slit prompt-gamma camera for locating of distal dose falloff in proton therapy”.Nuclear Engineering and Technology,51(5), 1406 (2019). doi:10.1016/j.net.2019.03.008
2019 doi
-
[14]
Analytical modeling and Monte Carlo simulations of multi-parallel slit and knife-edge slit prompt gamma cameras
B. F. Huismanet al.“Analytical modeling and Monte Carlo simulations of multi-parallel slit and knife-edge slit prompt gamma cameras”.Physics in Medicine and Biology, 68 (2023). doi:10.1088/1361-6560/acd237
2023 doi
-
[15]
Tackling range uncertainty in proton therapy: De- velopment and evaluation of a new multi-slit prompt-gamma camera (mspgc) system
Y. Ku et al. “Tackling range uncertainty in proton therapy: De- velopment and evaluation of a new multi-slit prompt-gamma camera (mspgc) system”. Nuclear Engineering and Technology, 55, 3140 (2023). doi:10.1016/j.net.2023.05.028
2023 doi
-
[16]
MO-FG-CAMPUS-JeP1-01: Prompt Gamma Imag- ing with a Multi-Knife-Edge Slit Collimator: Evaluation for Use in Pro- ton Beam Range Verification
J. Ready et al. “MO-FG-CAMPUS-JeP1-01: Prompt Gamma Imag- ing with a Multi-Knife-Edge Slit Collimator: Evaluation for Use in Pro- ton Beam Range Verification”. Medical Physics, 43(6), 3717 (2016). doi:10.1118/1.4957338
2016 doi
-
[17]
J. F. Ready.Development of a multi-knife-edge slit collimator for prompt gamma ray imaging during proton beam cancer therapy. PhD thesis, Uni- versity of California, Berkeley (2016).http://adsabs.harvard.edu/abs/ 2016PhDT.......151R
2016
-
[18]
Simulation Study of a 3-D Multislit Prompt Gamma Imaging System for Proton Therapy Monitoring
W. Lu et al. “Simulation Study of a 3-D Multislit Prompt Gamma Imaging System for Proton Therapy Monitoring”. IEEE Transac- tions on Radiation and Plasma Medical Sciences , 6, 355 (2022). doi:10.1109/TRPMS.2021.3107411
2022
-
[19]
Coded-aperture imaging systems: Past, present and future development – a review
M. J. Cieślak, K. A. Gamage, and R. Glover. “Coded-aperture imaging systems: Past, present and future development – a review”. Radiation Measurements, 92, 59 (2016). doi:10.1016/j.radmeas.2016.08.002. 25
2016 doi
-
[20]
Coded aperture imaging with uniformly redundant arrays
E. E. Fenimore and T. M. Cannon. “Coded aperture imaging with uniformly redundant arrays”. Applied Optics, 17, 337 (1978). doi:10.1364/AO.17.000337
1978 doi
-
[21]
New family of binary ar- rays for coded aperture imaging
S. R. Gottesman and E. E. Fenimore. “New family of binary ar- rays for coded aperture imaging”. Applied Optics, 28, 4344 (1989). doi:10.1364/ao.28.004344
1989 doi
-
[22]
Fourier transform photography: A new method for x-ray astronomy
J. G. Ables. “Fourier transform photography: A new method for x-ray astronomy”. Publications of the Astronomical Society of Australia, 1, 172 (1968). doi:10.1017/S1323358000011292
1968 doi
-
[23]
Scatter-hole cameras for x-rays and gamma rays
R. H. Dicke. “Scatter-hole cameras for x-rays and gamma rays”. ApJL, 153, L101 (1968). doi:10.1086/180230
1968 doi
-
[24]
Near-field high-resolution coded aperture gamma-ray imaging with separable masks
S. Sun, Y. Liu, and X. Ouyang. “Near-field high-resolution coded aperture gamma-ray imaging with separable masks”. Nuclear Instru- ments and Methods in Physics Research Section A: Accelerators, Spec- trometers, Detectors and Associated Equipment , 951, 163001 (2020). doi:10.10...
2020
-
[26]
The SiFi-CC project – Feasibility study of a scintillation- fiber-based Compton camera for proton therapy monitoring
J. Kasperet al.“The SiFi-CC project – Feasibility study of a scintillation- fiber-based Compton camera for proton therapy monitoring”.Physica Med- ica, 76, 317 (2020). doi:10.1016/j.ejmp.2020.07.013
2020 doi
-
[27]
Range monitoring capabilities with the sifi-cc de- tector: Spectral-spatial imaging with monte carlo-simulated data
J. Roser et al. “Range monitoring capabilities with the sifi-cc de- tector: Spectral-spatial imaging with monte carlo-simulated data”. Acta Physica Polonica B Proceedings Supplement , 17, 1 (2024). doi:10.5506/APhysPolBSupp.17.7-A9
2024 doi
- [28]
-
[29]
A systematic study of LYSO:Ce, LuAG:Ce and GAGG:Ce scintillating fibers properties
K. Rusiecka et al. “A systematic study of LYSO:Ce, LuAG:Ce and GAGG:Ce scintillating fibers properties”. Journal of Instrumentation, 16(11), P11006 (2021). doi:10.1088/1748-0221/16/11/P11006
2021 doi
-
[30]
Taiwan Applied Crystal webpage.https://www.tacrystal.com/
-
[31]
https://docs.broadcom
Broadcom AFBR-S4N44P164M Data sheet. https://docs.broadcom. com/doc/AFBR-S4N44P164M-DS-4x4-NUV-MT-Silicon-Photo-Multiplier-Array
-
[32]
Offset: Optical fiber folded scintillating extended tracker
D. Lo Presti et al. “Offset: Optical fiber folded scintillating extended tracker”. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 737, 195 (2014). doi:10.1016/j.nima.2013.11.049. 26
2014 doi
-
[33]
https://www.wacker
Data sheet of ELASTOSIL RT 604, WACKER. https://www.wacker. com/h/en-gb/medias/ELASTOSIL-RT-604-AB-en-2024.06.16.pdf
2024
-
[34]
TOFPET 2: A high-performance circuit for PET time-of-flight
A. Di Francescoet al. “TOFPET 2: A high-performance circuit for PET time-of-flight”. Nuclear Instruments and Methods in Physics Research Sec- tion A: Accelerators, Spectrometers, Detectors and Associated Equipment, 824, 194 (2016). doi:10.1016/j.nima.2015.11.036
2016 doi
-
[35]
SiPM Readout System
PETsys. “SiPM Readout System” (2023). https://www. petsyselectronics.com/web/website/docs/products/product4/ Long%20Flyer%20SiPM%20readout%20chain_v15.pdf
2023
-
[36]
Comparison of readout systems for high-rate silicon pho- tomultiplier applications
M. Wonget al. “Comparison of readout systems for high-rate silicon pho- tomultiplier applications”.Journal of Instrumentation, 19, P01019 (2024). doi:10.1088/1748-0221/19/01/P01019
2024 doi
-
[37]
X-Ray Mass Attenuation Coefficients
NIST. “X-Ray Mass Attenuation Coefficients”. https://physics.nist. gov/PhysRefData/XrayMassCoef/ElemTab/z74.html
-
[38]
Evaluation of fiber-shaped lyso for double readout gamma photon detection
P. Anfréet al. “Evaluation of fiber-shaped lyso for double readout gamma photon detection”.IEEE Transactions on Nuclear Science, 54, 391 (2007). doi:10.1109/TNS.2007.892942
2007
-
[39]
Stopping-power and range tables for protons
NIST. “Stopping-power and range tables for protons”.https://physics. nist.gov/PhysRefData/Star/Text/PSTAR-t.html
-
[40]
Geant4—a simulation toolkit
S. Agostinelli et al. “Geant4—a simulation toolkit”. Nuclear Instru- ments and Methods in Physics Research Section A: Accelerators, Spec- trometers, Detectors and Associated Equipment , 506(3), 250 (2003). doi:10.1016/S0168-9002(03)01368-8
2003 doi
-
[41]
Geant4: Physics List Guide
“Geant4: Physics List Guide”. https://geant4-userdoc.web.cern.ch/ UsersGuides/PhysicsListGuide/fo/PhysicsListGuide.pdf. Accessed: 2024-10-29
2024
-
[42]
Prompt-gamma emission in geant4 revisited and confronted with experiment
A. Wrońska et al. “Prompt-gamma emission in geant4 revisited and confronted with experiment”. Physica Medica, 88, 250 (2021). doi:https://doi.org/10.1016/j.ejmp.2021.07.018
2021 doi
-
[43]
Goddess: A geant4 extension for easy modelling of optical detector components
E. Dietz-Laursonnet al. “Goddess: A geant4 extension for easy modelling of optical detector components”.Journal of Instrumentation, 12, P04026 (2017). doi:10.1088/1748-0221/12/04/P04026
2017 doi
-
[44]
TOFPET2 ASIC SiPM Readout System - Soft- ware User Guide
P. Electronics. “TOFPET2 ASIC SiPM Readout System - Soft- ware User Guide” (2024). https://www.petsyselectronics.com/ web/website/documentation/TOFPET2%20Downloads/Documentation/ PETsys%20TOFPET%202%20ASIC%20SiPM%20Readout%20System%20-% 20Software%20User%20Guide%20v2024.08.pdf
2024
-
[45]
Root — an object oriented data analysis framework
R. Brun and F. Rademakers. “Root — an object oriented data analysis framework”. Nuclear Instruments and Methods in Physics Research Sec- tion A: Accelerators, Spectrometers, Detectors and Associated Equipment, 389(1), 81 (1997). doi:10.1016/S0168-9002(97)00048-X. 27
1997 doi
-
[46]
sifi-framework. A data analysis software
SiFi-CC. “sifi-framework. A data analysis software”. https://github. com/SiFi-CC/sifi-framework/tree/4to1_classes_HIT_refactoring
-
[47]
Maximum likelihood reconstruction for emis- sion tomography
L. A. Shepp and Y. Vardi. “Maximum likelihood reconstruction for emis- sion tomography”. IEEE Transactions on Medical Imaging, 1(2), 113–122 (1982). doi:10.1109/tmi.1982.4307558
1982
-
[48]
Em reconstruction algorithms for emission and transmission tomography
K. Lange and R. Carson. “Em reconstruction algorithms for emission and transmission tomography”. IEEE Transactions on Medical Imaging, 8(2), 306 (1984)
1984
-
[49]
Machine learning-based patient specific prompt-gamma dose monitoring in proton therapy
P. Gueth et al. “Machine learning-based patient specific prompt-gamma dose monitoring in proton therapy”. Physics in Medicine and Biology, 58(13), 4563–4577 (2013). doi:10.1088/0031-9155/58/13/4563
2013 doi
-
[50]
TOFPET2: a high-performance ASIC for time and amplitude measurements of SiPM signals in time-of-flight ap- plications
P. Electronics. “TOFPET2: a high-performance ASIC for time and amplitude measurements of SiPM signals in time-of-flight ap- plications”. https://www.petsyselectronics.com/web/website/docs/ products/product1/TOFPET2_Overview.pdf
-
[51]
Design and performance evaluation of a codedapertureimagingsystemforreal-timepromptgamma-raymonitoring during proton therapy
S. Sun, Y. Liu, and X. Ouyang. “Design and performance evaluation of a codedapertureimagingsystemforreal-timepromptgamma-raymonitoring during proton therapy”. Radiation Physics and Chemistry, 174 (2020). doi:10.1016/j.radphyschem.2020.108891
2020
-
[52]
Design optimisation of a TOF-based collimated camera prototype for online hadrontherapy monitoring
M. Pinto et al. “Design optimisation of a TOF-based collimated camera prototype for online hadrontherapy monitoring”.Physics in Medicine and Biology, 59(24), 7653 (2014). doi:10.1088/0031-9155/59/24/7653. 28
2014 doi
Reviewed August 10, 2026 · model on record in the stance chip above.
Discussion (0). Continue with ORCID to comment.