Characterization of a novel plastic scintillation detector for in vivo electron dosimetry
Pith reviewed 2026-05-18 19:08 UTC · model grok-4.3
The pith
A novel plastic scintillation detector delivers surface dose measurements in electron beams that agree with ionization chambers to 1.3% on average.
A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.
Core claim
The authors establish that their plastic scintillation detector is suitable for clinically acceptable electron beam dosimetry, as evidenced by its linearity, dose rate independence, isotropy, reproducibility with total variation below 2%, agreement in depth dose parameters R50 and R80 within 1 mm, lateral profiles with mean absolute error below 1.5%, small field output within 2%, and surface doses agreeing to 1.3% with ionization chambers and 2.1% with radiochromic film.
What carries the argument
Dual-channel Cherenkov radiation correction that separates the scintillation signal from the unwanted Cherenkov light generated in the fiber.
If this is right
- The PSD can provide real-time dosimetry during electron beam treatments.
- Surface dose measurements become possible with accuracy comparable to ionization chambers and film.
- The detector maintains performance across various field sizes and depths in electron beams.
- It offers high spatial resolution suitable for in vivo applications.
Where Pith is reading between the lines
- This approach may extend to other radiation types if Cherenkov correction is adapted.
- Integration with treatment machines could allow for adaptive radiotherapy based on real-time dose feedback.
- Future studies might test it in clinical patient scenarios to confirm in vivo performance.
Load-bearing premise
The dual-channel Cherenkov radiation correction accurately isolates the scintillation signal without residual artifacts or added uncertainty in the electron beam environment.
What would settle it
A surface dose measurement with the PSD that deviates by more than 5% from the ionization chamber result in a standard electron beam setup on a solid water phantom would contradict the claim of clinical acceptability.
Figures
read the original abstract
Introduction: Real-time dosimetry of surface doses in electron beams has not been widely established yet. Plastic scintillation detectors (PSD) promise high spatial resolution and real-time dosimetry with minimum perturbation of the radiation field. This study characterizes a novel PSD in an electron beam to determine its suitability for in vivo dosimetry. Methods: Dual-channel Cherenkov radiation correction and dosimetric characterization of the PSD were investigated using reference ionization chambers. Percentage depth-dose curves, lateral profiles, and output factors were compared with reference ionization chamber measurements. Surface doses were measured on solid water and on an anthropomorphic phantom and were compared to ionization chamber and radiochromic film measurements. Results: The investigated PSD demonstrated clinically acceptable linearity, dose rate independence, isotropy and reproducibility (total variation <2%). Dosimetric deviation in R50 and R80 were below 1.0 mm and lateral profiles agreed with a mean absolute error below 1.5%. Small field measurements were within 2% of the reference ionization chamber results. Surface dose measurements had mean relative deviations of 1.3% from ionization chamber measurements and 2.1% from radiochromic film measurements. Conclusion: The PSD investigated in this study is suitable for clinically acceptable electron beam dosimetry and provides accurate dosimetric results for surface dose measurements. It has the potential to be used for real-time in vivo dosimetry.
Editorial analysis
A structured set of objections, weighed in public.
Referee Report
Summary. The manuscript characterizes a novel plastic scintillation detector (PSD) for in vivo electron dosimetry. It investigates dual-channel Cherenkov radiation correction and compares percentage depth-dose curves, lateral profiles, output factors, and surface doses measured on solid water and anthropomorphic phantoms to reference ionization chamber and radiochromic film measurements. The PSD shows clinically acceptable performance with total variation <2%, deviations in R50/R80 below 1.0 mm, lateral profiles MAE <1.5%, small field within 2%, and surface dose mean deviations of 1.3% vs chamber and 2.1% vs film, concluding suitability for real-time in vivo dosimetry.
Significance. Should the results be confirmed with more detailed uncertainty analysis, this PSD could fill an important niche for real-time surface dose monitoring in electron beam radiotherapy, where such dosimetry is not widely established. The use of independent reference standards for validation is a strength, supporting potential clinical adoption for high-resolution, low-perturbation measurements.
major comments (1)
- [Methods] Methods: The dual-channel Cherenkov radiation correction is central to isolating the scintillation signal for dosimetric characterization and surface dose comparisons. The manuscript does not quantify post-correction Cherenkov leakage or residual bias, despite variations in Cherenkov yield with depth, field size, and phantom material in electron beams. This is load-bearing for the reported surface dose accuracy (mean relative deviations of 1.3% vs. ionization chamber and 2.1% vs. film) and the claim of total variation <2%.
minor comments (2)
- [Abstract] Abstract: The abstract notes small deviations but lacks full details on setup, uncertainty budgets, and potential confounders such as temperature or cable effects; these should be expanded in the main text for transparency.
- [Results] Results: Adding explicit uncertainty estimates or error bars to the reported mean deviations would strengthen assessment of the clinically acceptable performance claims.
Simulated Author's Rebuttal
We thank the referee for the constructive review and the recommendation for major revision. We have addressed the single major comment by agreeing that additional quantification is warranted and committing to revisions that strengthen the description of the Cherenkov correction without altering the original experimental data or conclusions.
read point-by-point responses
-
Referee: [Methods] Methods: The dual-channel Cherenkov radiation correction is central to isolating the scintillation signal for dosimetric characterization and surface dose comparisons. The manuscript does not quantify post-correction Cherenkov leakage or residual bias, despite variations in Cherenkov yield with depth, field size, and phantom material in electron beams. This is load-bearing for the reported surface dose accuracy (mean relative deviations of 1.3% vs. ionization chamber and 2.1% vs. film) and the claim of total variation <2%.
Authors: We agree with the referee that an explicit quantification of residual Cherenkov leakage after dual-channel correction would strengthen the manuscript, given that Cherenkov yield can vary with depth, field size, and phantom material. The original manuscript applies the standard dual-channel correction but does not report a direct estimate of post-correction residual bias. The observed agreement with independent reference detectors (1.3% mean deviation versus ionization chamber and 2.1% versus film) across multiple conditions provides indirect support that any residual is small enough to remain within the stated clinical tolerances. To address the comment directly, we will revise the Methods section to describe how residual bias is estimated from the channel difference in low-scintillation reference conditions and will add corresponding results confirming that the residual remains below the level that would affect the reported total variation or surface-dose accuracies. revision: yes
Circularity Check
No significant circularity: experimental validation against independent references
full rationale
This is an experimental characterization paper that measures PSD performance (linearity, isotropy, surface dose) by direct comparison to external reference standards including ionization chambers and radiochromic film. No first-principles derivations, fitted parameters renamed as predictions, or self-citation chains appear in the provided abstract or described methods. The dual-channel Cherenkov correction is a procedural step whose accuracy is assessed via the same independent benchmarks rather than being defined into the result. The study is therefore self-contained against external benchmarks and receives a non-circularity finding.
Axiom & Free-Parameter Ledger
axioms (1)
- domain assumption Reference ionization chambers and radiochromic film provide accurate ground-truth dose measurements in the electron beam setups used.
Reference graph
Works this paper leans on
-
[1]
Review of electron beam therapy physics
Hogstrom KR, Almond PR. Review of electron beam therapy physics. Phys Med Biol. 2006;51(13):R455-
work page 2006
-
[2]
doi:10.1088/0031-9155/51/13/R25
-
[3]
Wolfe CM, Green WH, Hatfield HK, Shakar TJ, Baniahmad O, Cognetta AB. Multiple secondary cutane ous tumours following electron beam radiotherapy for cutaneous malignancies of the scalp. Australas J Dermatol. 2012;53(3):233-238. doi:10.1111/j.1440-0960.2012.00917.x
-
[4]
Quantitative approaches in electron skin collimation for the practical benefits
Wang D, Polignani JA. Quantitative approaches in electron skin collimation for the practical benefits. J Appl Clin Med Phys. 2024;25(4):e14236. doi:10.1002/acm2.14236
-
[5]
Gerbi BJ, Antolak JA, Deibel FC, et al. Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25. Med Phys. 2009;36(7):3239-3279. doi:10.1118/1.3125820
-
[6]
Gafchromic EBT3 film dosimetry in electron be ams - energy dependence and improved film read -out
Sipilä P, Ojala J, Kaijaluoto S, Jokelainen I, Kosunen A. Gafchromic EBT3 film dosimetry in electron be ams - energy dependence and improved film read -out. J Appl Clin Med Phys. 2016;17(1):360 -373. doi:10.1120/jacmp.v17i1.5970
-
[7]
Bufacchi A, Carosi A, Adorante N, et al
18. Bufacchi A, Carosi A, Adorante N, et al. In vivo EBT radiochromic film dosimetry of electron beam for Total Skin Electron Therapy (TSET). Phys Med. 2007;23(2):67-72. doi:10.1016/j.ejmp.2007.03.003
-
[8]
Characterization of a new radiochromic film (LD-V1) using mammographic be am qualities
Nakajima E, Sato H. Characterization of a new radiochromic film (LD-V1) using mammographic be am qualities. Z Med Phys. 2025;35(2):169-176. doi:10.1016/j.zemedi.2023.05.004
-
[9]
Small field electron beam dosimetry using MOSFET detector
Amin MN, Heaton R, Norrlinger B, Islam MK. Small field electron beam dosimetry using MOSFET detector. J Appl Clin Med Phys. 2010;12(1):3267. doi:10.1120/jacmp.v12i1.3267
-
[10]
Ciocca M, Piazzi V, Lazzari R, et al. Real -time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early -stage breast cancer. Radiother Oncol. 2006;78(2):213-216. doi:10.1016/j.radonc.2005.11.011
-
[11]
Characterization of the response of commercial diode detectors use d for in vivo dosimetry
Meiler RJ, Podgorsak MB. Characterization of the response of commercial diode detectors use d for in vivo dosimetry. Med Dosim. 1997;22(1):31-37. doi:10.1016/s0958-3947(96)00152-5
-
[12]
Ade N, Du Plessis FCP. Electron beam dose perturbations caused by diode detectors used for in vivo dosimetry: Gafchromic film dose measurements in a re alistic pelvic prosthesis phantom. Radiation Physics and Chemistry. 2018;151:232-238
work page 2018
-
[14]
Water-equivalent plastic scintillation detectors for high-energy be am dosimetry: I
Beddar AS, Mackie TR, Attix FH. Water-equivalent plastic scintillation detectors for high-energy be am dosimetry: I. Physical characteristics and theoretical consideration. Phys Med Biol. 1992;37(10):1883-
work page 1992
-
[15]
doi:10.1088/0031-9155/37/10/006
-
[17]
Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy
Beaulieu L, Beddar S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Phys Med Biol. 2016;61(20):R305-R343. doi:10.1088/0031-9155/61/20/R305
-
[18]
Dimitriadis A, Patallo IS, Billas I, Duane S, Nisbet A, Clark CH. Characterisation of a plastic scintillation detector to be used in a multicentre stereotactic radiosurgery dosimetry audit. Radiat Phys Che m. 2017;140:373-378
work page 2017
-
[19]
Plastic scintillator -based dosimeters for ultra-high dose rate (UHDR) electron radiotherapy
Ciarrocchi E, Ravera E, Cavalieri A, et al. Plastic scintillator -based dosimeters for ultra-high dose rate (UHDR) electron radiotherapy. Phys Med. 2024;121:103360. doi:10.1016/j.ejmp.2024.103360
-
[20]
Vanreusel V, Gasparini A, Galante F, et al. Point scintillator dosimetry in ultra-high dose rate ele ctron “FLASH” radiation therapy: A first characterization. Phys Med. 20 22;103:127-137. doi:10.1016/j.ejmp.2022.10.005
-
[21]
Liu K, Holmes S, Schüler E, Beddar S. A comprehensive investigation of the performance of a comme rcial scintillator system for applications in electron FLASH radiotherapy. Med Phys. 2024;51(6):4504-4512. doi:10.1002/mp.17030
-
[22]
Baikalov A, Tho D, Liu K, Bartzsch S, Beddar S, Schüler E. Characterization of a Time-Resolved, Real-Time Scintillation Dosimetry System for Ultra-High Dose-Rate Radiation Therapy Applications. Int J Radiat Oncol Biol Phys. 2025;121(5):1372-1383. doi:10.1016/j.ijrobp.2024.11.092 15
-
[23]
Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams
Clift MA, Sutton RA, Webb DV. Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams. Phys Med Biol. 2000;45(5):1165 -1182. doi:10.1088/0031- 9155/45/5/307
-
[24]
Plastic scintillation dosimetry: optimization of light collection efficiency
Beddar AS, Law S, Suchowerska N, Mackie TR. Plastic scintillation dosimetry: optimization of light collection efficiency. Phys Med Biol. 2003;48(9):1141-1152. doi:10.1088/0031-9155/48/9/305
-
[25]
Blue Physics model 11: instructions for use
Blue Physics LLC. Blue Physics model 11: instructions for use. Version 1.1. 2023
work page 2023
-
[26]
Characteristics of a plastic scintillation detector in photon beam dosimetry
Das IJ, Sohn JJ, Lim SN, Sengupta B, Feijoo M, Yadav P. Characteristics of a plastic scintillation detector in photon beam dosimetry. J Appl Clin Med Phys. 2024;25(1):e14209. doi:10.1002/acm2.14209
-
[27]
Jacqmin DJ, Miller JR, Barraclough BA, Labby ZE. Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields. J Appl Clin Med Phys. 2022;23(8):e13728. doi:10.1002/acm2.13728
-
[28]
AAPM’s TG-51 protocol for clinical reference dosimetry of high- energy photon and electron beams
Almond PR, Biggs PJ, Coursey BM, et al. AAPM’s TG-51 protocol for clinical reference dosimetry of high- energy photon and electron beams. Med Phys. 1999;26(9):1847-1870. doi:10.1118/1.598691
-
[29]
Rapid and reversible adaptation of a clinical linear accelerator for electron FLASH radiotherapy
Schneider F, Bauer CJ, Göbel ID, et al. Rapid and reversible adaptation of a clinical linear accelerator for electron FLASH radiotherapy. Phys Med. 2025;136:105032. doi:10.1016/j.ejmp.2025.105032
-
[30]
Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR-linac
Oolbekkink S, van Asselen B, Woodings SJ, et al. Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR-linac. J Appl Clin Med Phys. 2024;25(1):e14180. doi:10.1002/acm2.14180
-
[31]
Dosimetric characterization of a novel commercial plastic scintillation detector with an MR-Linac
Ferrer C, Hue rtas C, García D, Sáez M. Dosimetric characterization of a novel commercial plastic scintillation detector with an MR-Linac. Med Phys. 2023;50(4):2525-2539. doi:10.1002/mp.16204
-
[32]
Task Group 142 report: quality assurance of medical accelerators
Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197-4212. doi:10.1118/1.3190392
-
[33]
Characterization of the plastic scintillation detector Exradin W2 for small field dosimetry
Galavis PE, Hu L, Holmes S, Das IJ. Characterization of the plastic scintillation detector Exradin W2 for small field dosimetry. Med Phys. 2019;46(5):2468-2476. doi:10.1002/mp.13501
-
[34]
Characterization of the Exradin W1 scintillator for use in radiothe rapy
Carrasco P, Jornet N, Jordi O, et al. Characterization of the Exradin W1 scintillator for use in radiothe rapy. Med Phys. 2015;42(1):297-304. doi:10.1118/1.4903757
-
[35]
Post-mastectomy radiotherapy: Impact of bolus thickness and irradiation technique on skin dose
Miéville FA, Pitteloud N, Achard V, et al. Post-mastectomy radiotherapy: Impact of bolus thickness and irradiation technique on skin dose. Z Med Phys. 2024;34(4):542-554. doi:10.1016/j.zemedi.2023.03.004
-
[36]
Effect of bolus materials on dose deposition in deep tissues during ele ctron beam radiotherapy
Kong D, Wu J, Kong X, et al. Effect of bolus materials on dose deposition in deep tissues during ele ctron beam radiotherapy. J Radiat Res. 2024;65(2):215-222. doi:10.1093/jrr/rrae001
-
[37]
Electron beam characteristics at extende d source -to-surface distances for irregular cut-outs
Arunkumar T, Supe SS, Ravikumar M, Sathiyan S, Ganesh M. Electron beam characteristics at extende d source -to-surface distances for irregular cut-outs. J Med Phys. 2010;35(4):207-214. doi:10.4103/0971- 6203.71763
-
[38]
Schulz JB, Gibson C, Dubrowski P, et al. Shaping success: clinical implementation of a 3D-printed ele ctron cutout program in external beam radiation therapy. Front Oncol. 2023;13:1237037 . doi:10.3389/fonc.2023.1237037
-
[39]
Field output correction factors using a scintillation detector
Gingras L, Cervantes Y, Beaulieu F, et al. Field output correction factors using a scintillation detector. Med Phys. 2025;52(6):4844-4861. doi:10.1002/mp.17729
discussion (0)
Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.