MOTION, a liquid xenon time projection chamber platform for high voltage technologies in dark matter detectors

T0 review · 3 major / 5 minor · reviewed 2026-08-06 · deepseek-v4-flash

Pith's one-line read A 70 kg liquid xenon detector platform sets out to explain why high-voltage systems in dark matter TPCs break down well below theoretical limits, by measuring dielectric breakdown in controlled conditions up to 200 kV.

desk verdict A well-executed commissioning report for a new 70 kg LXe HV test platform; the facility is real and worth knowing about, but the headline claim about enabling controlled 200 kV breakdown studies is a plan, not yet a demonstrated capability. read the letter →

arxiv 2608.04679 v1 pith:67KIISYR submitted 2026-08-05 physics.ins-det

classification physics.ins-det
keywords liquidxenontimeprojectionchamberhighvoltagedielectricbreakdowndarkmatterdetectionpurificationpuritymonitorelectrodesurface
topics Dark Matter
verification ladder T0 review T1 audit T2 compute T3 formal

The pith

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The reading

The paper presents MOTION, a 70 kg liquid xenon (LXe) detector built to study the high-voltage hardware inside the liquid xenon time projection chambers used in dark matter searches. Its central claim is that this facility, with electrodes biased up to 200 kV negative polarity inside a cryostat, provides a dedicated platform for systematic studies of dielectric breakdown, pre-breakdown currents, field emission, and discharge phenomena in LXe. The motivation is a known operational problem: tonne-scale LXe TPCs have repeatedly failed to reach their design drift fields, breaking down at fields as low as -50 kV/cm even though the bulk dielectric strength of LXe is near 1 MV/cm, with surface asperities and stressed electrode area suspected as the cause. MOTION is equipped to test this by correlating optical, electrical, and purity measurements with controlled electrode geometry and surface finish. If the platform works as claimed, it would supply the breakdown statistics and component-validation data needed to scale reliably to the next-generation 60-80 tonne liquid xenon observatory.

What carries the argument

The central object is the electrode assembly: a negatively biased cathode (60 mm diameter) and a grounded anode (80 mm diameter), both with Rogowski profiles (electrode contours shaped to give a uniform electric field between parallel disks), separated by 10 mm and immersed in LXe, with cathode bias up to -200 kV. The diagnostics carry the experimental argument: a transimpedance amplifier on the anode reads pre-breakdown currents; a camera through a viewport observes the gap for electroluminescence, micro-arcs, and bubble dynamics; two VUV-sensitive SiPMs monitor scintillation and breakdown precursors; and a purity monitor measures electron lifetime to quantify electronegative impurities. Around this core, a surface-characterization protocol using confocal laser microscopy maps areal roughness parameters ($S_a$, $S_z$, $S_{sk}$, $S_{ku}$) over entire electrodes. The component under validation is a compression-fitting HV feedthrough with conductive-polyethylene field grading and a UHMWPE insulator.

What would settle it

After the purity monitor is installed, repeat the same cathode ramp on identical electrode pairs differing only in surface finish, with electron lifetime held stable: if the polished pair does not break down at a higher voltage or lower rate than the lathed pair, the surface-asperity picture that motivates MOTION is falsified; if electron lifetime drifts between runs, the controlled-condition premise fails.

Watch

Extended reading notes

Core claim

The paper's central claim is that MOTION closes a gap in current LXe TPC development: no existing facility offers a controlled, instrumented liquid-xenon environment dedicated to high-voltage component behaviour at this voltage scale. The detector consists of two Rogowski-profile electrodes immersed in LXe, a negatively biased cathode facing a grounded anode across a 10 mm gap, with the cathode voltage rampable to -200 kV. Its purpose is to measure how dielectric breakdown depends on electrode geometry, surface condition, applied voltage, and electronegative impurity level, and to correlate discharge precursors with breakdown events through simultaneous pre-breakdown current readout, camera imaging, and SiPM light detection. The paper also presents a custom HV feedthrough design for testing at full scale and a laser-microscopy protocol for quantifying electrode surface topography before and after treatment. Commissioning results reported include xenon transfer, recirculation up to 10 SLPM, temperature stability at 175 K, and recuperation; the SiPM and purity-monitor upgrades are ongoing.

Load-bearing premise

That the liquid xenon purity in the inner vessel will be high enough for the planned high-voltage measurements to be meaningful; the paper reports no electron-lifetime measurement yet, and if electronegative impurities sit at levels that distort drift or breakdown probability, the platform cannot deliver the controlled breakdown data it is built for.

Editorial extensions

If this is right

  • Breakdown data from controlled ramps will quantify the probability distribution of discharge voltage, not just a threshold, giving survival probabilities for HV components in large TPCs.
  • If surface finish is the dominant lever, the laser-microscopy quality-assurance protocol will identify which polishing and passivation treatments are worth applying to future large-scale electrodes, even where full-surface scanning of 3 m components is not feasible.
  • The purity monitor will test the hypothesis that electronegative impurities in LXe suppress discharges, which would force a trade-off between charge-drift lifetime and HV stability in future detectors.
  • Successful validation of the feedthrough design in LXe at realistic voltage would de-risk the high-voltage delivery chain for the next-generation observatory.
  • Simultaneous optical and electrical detection of discharge precursors may provide an early-warning signature usable in operating TPCs to mitigate spurious electron backgrounds.

Reading between the lines

Editorial extensions of the paper, not claims the author makes directly.

  • Editorial inference: if surface asperities are the controlling factor, MOTION's full-electrode topographic maps could be used to build a breakdown-risk map that predicts discharge sites from local $S_z$ anomalies; the paper describes the maps but does not itself make that prediction.
  • Editorial inference: the measured 10 SLPM recirculation ceiling sets an upper bound on achievable xenon purity, so if impurities suppress breakdown the facility may need to map a purity-breakdown trade-off grid, a study the paper lists as planned but does not yet constrain.
  • Editorial inference: the same high-voltage dielectric questions apply to liquid argon TPCs and to neutrinoless double beta decay searches, so the systematic protocol could transfer beyond dark matter WIMP detectors.
  • Editorial inference: a direct test of stressed-electrode-area scaling would vary electrode diameter and gap independently, and MOTION's adjustable anode position provides a route, though no such data are shown yet.
Share X Bluesky LinkedIn Reddit HN

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, and a circularity audit.

Referee Report

3 major / 5 minor

Summary. The paper reports the construction and commissioning of MOTION, a 70 kg liquid xenon facility at KIT intended as a test platform for high-voltage technologies, dielectric-breakdown studies, and HV feedthrough development for the XLZD observatory. Sections 2–5 describe the gas handling, cryogenic system, cryostat, instrumentation, and slow control. Section 6 presents commissioning results: xenon transfer, recirculation at up to 10 SLPM (limited by compressor outlet pressure), a 16 h temperature stability run with mean 175.34 ± 0.53 K, and recuperation. Section 7 develops a confocal-laser-microscopy protocol for electrode surface roughness, and Section 8 describes ongoing upgrades (SiPM characterization, a purity monitor) and planned measurements, including 200 kV breakdown tests and HV feedthrough validation. The abstract claims that MOTION 'enables controlled studies of dielectric breakdown in LXe' up to 200 kV.

Significance. If the facility performs as claimed, it would fill a clear R&D gap for XLZD: systematic, controlled measurements of dielectric breakdown in liquid xenon at up to 200 kV, with optical and electrical diagnostics and quantified electrode surface conditions. The paper's strengths are its detailed engineering documentation and the direct commissioning data—leak rate 2e-7 mbar L/s, 16 h temperature stability at 175.34 K, maximum recirculation of 10 SLPM, and a reproducible surface-roughness protocol with ISO-filtered areal parameters. These are concrete, useful contributions. The central caveat is that the 'controlled studies' capability is currently conditional on a purity monitor that is described as an ongoing upgrade, and no electron-lifetime or impurity measurement is reported anywhere in the commissioning results.

major comments (3)
  1. [Abstract; Section 8.2] The abstract and Section 1 state that MOTION 'enables controlled studies of dielectric breakdown in LXe' and is 'a unique platform for systematic studies,' but no electron-lifetime or impurity measurement is reported in Section 6, and Section 8.2 describes the purity monitor as an ongoing upgrade. Section 8.2 also notes, citing XeBrA, that electronegative impurities 'may also decrease the probability of high-voltage breakdown.' Without a measured and preferably controlled impurity concentration, breakdown-voltage data from the planned 200 kV tests cannot be interpreted as a function of electrode geometry, surface condition, and applied voltage alone. The manuscript should either report an electron-lifetime measurement from the commissioning period or re-scope the capability claims (e.g., 'designed to enable') and identify the purity monitor as a prerequisite for the systematic studies.
  2. [Section 6.2] The commissioning establishes 10 SLPM as the maximum recirculation flow because the compressor outlet reaches 2.9 bar at its 3 bar rating, but it does not demonstrate that this flow rate achieves the purity required for valid HV studies. No electron lifetime or impurity concentration is quoted after getter recirculation, so the link between recirculation flow and the environmental variable that Section 8.2 identifies as affecting breakdown probability is missing. Please add a quantitative purity target and, ideally, a measurement showing that the 10 SLPM recirculation loop meets that target.
  3. [Sections 4.2 and 8.3] No high-voltage operation of the 200 kV chain is reported: the only electrical results are bench characterization of the transimpedance amplifier (Section 4.2), while the TVS protection board 'will be implemented' and all HV-breakdown measurements are in Section 8.3 'Planned measurements.' For a construction and commissioning paper this is acceptable, but the abstract's present-tense claim that MOTION enables controlled up-to-200-kV studies exceeds the demonstrated state; the wording should be changed to reflect that HV operation is the subject of ongoing and planned work.
minor comments (5)
  1. [Section 6.3] The quoted ±0.53 K is the systematic measurement uncertainty, not the observed stability; please also report the standard deviation or range of the 16 h temperature series, since the plotted spread appears to exceed 0.53 K.
  2. [Section 4.1] The level probe is characterized in liquid nitrogen, and the text states that in-situ calibration is still required; please state whether the commissioning fill used a calibrated level measurement to confirm that the electrodes were submerged.
  3. [Section 4.2] Please indicate whether the TIA linearity and bandwidth measurements included the TVS protection board or were performed without it, so that readers know which configuration is characterized.
  4. [Section 6.1] The integral leak rate is given as 2e-7 mbar L/s without specifying the test gas, method, or volume over which it was measured; adding this information would improve reproducibility.
  5. [Section 7] The full-surface scans require about 3 days at 20x and more than 15 days at 50x; a brief statement on how thermal drift and vibration are controlled over these long acquisition times would strengthen the protocol.

Circularity Check

0 steps flagged · score 0.0 of 10

No significant circularity: MOTION is a hardware construction and commissioning report whose central capability claim rests on direct engineering data, not on fitted targets or load-bearing self-citations.

full rationale

The paper contains no derivation chain that reduces to its own inputs. Its central claim—that MOTION is a 70 kg LXe platform for controlled high-voltage breakdown studies—is supported by physical construction details, commissioning measurements (xenon transfer, recirculation up to 10 SLPM limited by compressor outlet pressure, temperature stability at (175.34±0.53) K, recuperation), and instrumentation descriptions, rather than by a fitted parameter renamed as a prediction. The only fitted quantity, the level-probe linear fit in Section 4.1 and Figure 5, is a calibration of a capacitive level sensor and is explicitly qualified by the statement that 'an in-situ calibration is nonetheless required after installation'; it does not feed any physics result. The one self-citation, [49] (Xenoscope, with co-author Biondi), is used only as provenance for the purity-monitor readout boards and is not load-bearing for the facility's existence or for any predicted breakdown behavior. The paper's own stated caveat is a readiness/validity concern rather than a circularity: Section 8.2 notes electronegative impurities 'may also decrease the probability of high-voltage breakdown,' and the purity monitor is described as part of the ongoing upgrade, so the headline 'controlled studies' capability is conditional on measurements not yet reported. Conditionality and missing verification are not circularity, however. The design choices for the electrodes, condenser, and HV feedthrough are anchored to external references ([21], [22], [50]–[53]), and the commissioning data are reported as direct operational evidence. The paper is therefore self-contained as an instrumentation and commissioning report, and no circular step can be exhibited.

Assumptions & free parameters 2 free parameters · 3 assumptions · 0 invented entities

The central claims rest on well-known cryogenics and HV engineering practices, not on new physics entities. The only fitted numbers are calibration constants. No invented entities.

free parameters (2)
  • Level probe calibration line = slope -0.0296 pF/s, intercept 431.85 pF
    Linear fit to LN2 evaporation data (Fig 5). It calibrates the level probe but is not used in the paper's central capability claim.
  • Cold head setpoint temperature = 176 K
    Chosen operating point for temperature stability test; not fitted but selected by hand.
assumptions (3)
  • domain assumption Rogowski profile electrodes yield a uniform electric field between them
    Invoked without computation in Section 4, citing [22]; field uniformity is essential for interpreting breakdown measurements.
  • domain assumption Bulk dielectric strength of LXe is near 1 MV/cm but experiments observe lower values due to surface effects
    Stated in Section 1 as motivation, based on references [16,18,19]; the planned studies rely on this prior result.
  • standard math LN2 boiling at constant pressure provides a constant temperature cold source, and the GN2 buffer thermosiphon keeps the lower heat exchanger near 174 K
    Design assumption used in Section 3.5; simulations/testing estimate total cooling power around 130 W, but no data shown.

how reviews work

0 comments
Cite this review

Pith. "Pith review of MOTION, a liquid xenon time projection chamber platform for high voltage technologies in dark matter detectors." pith.science (2026). https://pith.science/paper/67KIISYR

@misc{pith2026260804679,
  author       = {Pith},
  title        = {Pith review of: MOTION, a liquid xenon time projection chamber platform for high voltage technologies in dark matter detectors},
  year         = {2026},
  howpublished = {\url{https://pith.science/paper/67KIISYR}},
  note         = {Machine review of arXiv:2608.04679}
}
read the original abstract

The XLZD observatory is a next-generation experiment designed to search for weakly interacting massive particles (WIMPs) and other rare events using a 60-80 tonne liquid xenon time projection chamber (TPC). This detector aims to achieve sensitivity across the full WIMP parameter space down to the neutrino fog, establishing the ultimate sensitivity for this dark matter search paradigm. This unprecedented scale introduces substantial engineering challenges and pushes operation into largely unexplored regimes: the interplay between high-voltage (HV) systems, liquid xenon, and conducting materials in ultra-pure environments. To systematically investigate these challenges, we have built MOTION, a 70 kg LXe detector dedicated to understanding HV performance and electrostatic phenomena up to 200 kV (negative polarity). We describe the design and construction of the experimental infrastructure, including the cryogenic system, xenon purification and storage. MOTION enables controlled studies of dielectric breakdown in LXe, permitting systematic characterization of discharge mechanisms and their dependence on electrode geometry, surface condition, and applied voltage. The detector also facilitates investigations of field emission and photoemission from electrodes following various surface treatments, and provides a platform for validating the design of an HV feedthrough constructed from radiopure materials. The insights from these studies are essential for ensuring the operational stability, radiopurity, and scalability required for next-generation dark matter detectors.

Discussion (0). Continue with ORCID to comment.

Reference graph

Works this paper leans on

39 extracted references · 32 canonical work pages

  1. [6]

    Baudis,Dark matter direct detection: status, results and future plans,J

    L. Baudis,Dark matter direct detection: status, results and future plans,J. Phys. Conf. Ser.3162 (2025) 012007 [2512.23039]. [7]XLZDcollaboration,The XLZD Design Book: towards the next-generation liquid xenon observatory for dark matter and neutrino physics,Eur. Phys. J. C85(2025) 1192 [2410.17137]

  2. [8]

    Aalbers et al.,A next-generation liquid xenon observatory for dark matter and neutrino physics,J

    J. Aalbers et al.,A next-generation liquid xenon observatory for dark matter and neutrino physics,J. Phys. G50(2023) 013001 [2203.02309]. [9]XLZDcollaboration,Neutrinoless double beta decay sensitivity of the XLZD rare event observatory, J. Phys. G52(2025) 045102 [2410.19016]. [10]XENONcollaboration,The XENONnT dark matter experiment,Eur. Phys. J. C84(202...

  3. [13]

    Malter,Anomalous Secondary Electron Emission A New Phenomenon,Physical Review49(1936) 478

    L. Malter,Anomalous Secondary Electron Emission A New Phenomenon,Physical Review49(1936) 478

  4. [14]

    Correlated Single- and Few-Electron Backgrounds Milliseconds after Interactions in Dual-Phase Liquid Xenon Time Projection Chambers

    A. Kopec, A.L. Baxter, M. Clark, R.F. Lang, S. Li, J. Qin et al.,Correlated single- and few-electron backgrounds milliseconds after interactions in dual-phase liquid xenon time projection chambers, JINST16(2021) [2103.05077]

  5. [15]

    Study and mitigation of spurious electron emission from cathodic wires in noble liquid time projection chambers

    A. Tomás, H.M. Araújo, A.J. Bailey, A. Bayer, E. Chen, B. López Paredes et al.,Study and mitigation of spurious electron emission from cathodic wires in noble liquid time projection chambers, Astropart. Phys.103(2018) 49 [1801.07231]

  6. [16]

    Watson, I

    J. Watson, I. Olcina, J. Soria, D.N. McKinsey, S. Kravitz, E.E. Deck et al.,Study of dielectric breakdown in liquid xenon with XeBrA: The xenon breakdown apparatus,Review of Scientific Instruments94(2023) 015112

  7. [17]

    Kuffel and P

    J. Kuffel and P. Kuffel,High voltage engineering fundamentals, Elsevier (2000)

  8. [18]

    Tvrznikova, E

    L. Tvrznikova, E. Bernard, S. Kravitz, K. O’Sullivan, G. Richardson, Q. Riffard et al.,Direct comparison of high voltage breakdown measurements in liquid argon and liquid xenon,JINST14 (2019) P12018

Show all 39 references
  1. [19]

    Massarczyk, P

    R. Massarczyk, P. Chu, C. Dugger, S. Elliott, K. Rielage and W. Xu,Paschen’s law studies in cold gases,JINST12(2017) P06019

  2. [20]

    Linstrom and W

    P. Linstrom and W. Mallard,NIST Chemistry WebBook, NIST Standard Reference Database, Number 69,https://doi.org/10.18434/T4D303, (accessed July 20, 2026)

  3. [21]

    Brown and et al.,PANCAKE: a large-diameter cryogenic test platform with a flat floor for next generation multi-tonne liquid xenon detectors,JINST19(2024) P05018

    A. Brown and et al.,PANCAKE: a large-diameter cryogenic test platform with a flat floor for next generation multi-tonne liquid xenon detectors,JINST19(2024) P05018

  4. [22]

    Trinh,Electrode Design for Testing in Uniform Field Gaps,IEEE Transactions on Power Apparatus and SystemsPAS-99(1980) 1235

    N. Trinh,Electrode Design for Testing in Uniform Field Gaps,IEEE Transactions on Power Apparatus and SystemsPAS-99(1980) 1235

  5. [23]

    Franzke,Microdischarges for analytical applications,Analytical and Bioanalytical Chemistry395 (2009) 547

    J. Franzke,Microdischarges for analytical applications,Analytical and Bioanalytical Chemistry395 (2009) 547

  6. [24]

    Smartec,Smartec UTI Sensor to Digital Transducer, DIL,https://www.mmselectronics.co.uk/ products/sensors/uti-transducer/smartec-uti-sensor-to-digital-transducer-dil, (accessed May 20, 2026). – 23 –

  7. [25]

    Analog Devices,AD7746 Datasheet and Product Info | Analog Devices, https://www.analog.com/en/products/ad7746.html, (accessed May 20, 2026)

  8. [26]

    FEMTO,DLPCA-200 switchable transimpedance amplifier, https://www.femto.de/en/current-transimpedance-amplifiers/ dc-up-to-500-khz-low-noise-variable-amplification-dlpca/, (accessed May 20, 2026). [27]DUNEcollaboration,Operation and performance of ProtoDUNE Dual Phase liquid arg...

  9. [28]

    National Instruments,What Is LabVIEW?,https://www.ni.com/en/shop/labview.html, (accessed July 29, 2026)

  10. [29]

    InfluxData Inc,InfluxData,https://www.influxdata.com/, (accessed July 29, 2026)

  11. [30]

    Grafana Labs,Full-stack observability for the agentic era | Grafana Labs, https://grafana.com/index/, (accessed July 29, 2026)

  12. [31]

    National Instruments,CompactRIO Systems (cRIO), https://www.ni.com/en/shop/compactrio.html, (accessed July 29, 2026)

  13. [32]

    KNF Group,N 630 serie ex-pumpen in atex-version, https://knf.com/de/de/loesungen/pumpen/serie/membran-gaspumpe-n-630-ex, (accessed July 24, 2026)

  14. [33]

    KEYENCE,Keyence 3d laser scanning microscope,https://www.keyence.de/products/ microscope/laser-microscope/vk-x3000/models/vk-x3050/, (accessed May 20, 2026)

  15. [34]

    P. A., M.H. Müser, N. Miller and et al.,The Surface-Topography Challenge: A Multi-Laboratory Benchmark Study to Advance the Characterization of Topography,Tribology Letters73(2025) 110

  16. [35]

    KEYENCE,Evaluation and filtering | surface roughness parameters according to din en iso 25178 | introduction to roughness measurement | keyence germany,https://www.keyence.de/ss/ products/microscope/roughness/surface/evaluation_process.jsp, (accessed May 20, 2026)

  17. [36]

    Digital Surf,Areal field parameters - surface metrology guide - digital surf, https://guide.digitalsurf.com/en/guide-areal-field-parameters.html, (Accessed 20 May 2026)

  18. [37]

    Wu,Electrode Realisation and SiPM Characterisation for Dark Matter Search with XENONnT and XLZD, Ph.D

    V.H.-S. Wu,Electrode Realisation and SiPM Characterisation for Dark Matter Search with XENONnT and XLZD, Ph.D. thesis, Karlsruhe Institute of Technology, 2026. 10.5445/IR/1000192347

  19. [38]

    Baudis, M

    L. Baudis, M. Galloway, A. Kish, C. Marentini and J. Wulf,Characterisation of Silicon Photomultipliers for liquid xenon detectors,JINST13(2018) P10022

  20. [39]

    Arneodo, M

    F. Arneodo, M. Benabderrahmane, G. Bruno, V. Conicella, A. Di Giovanni, O. Fawwaz et al., Cryogenic readout for multiple vuv4 multi-pixel photon counters in liquid xenon,Nucl. Instrum. Meth. A893(2018) 117

  21. [40]

    Peres,Advancing Multi-Messenger Astrophysics and Dark Matter Searches with XENONnT and the Top SiPM Array of Xenoscope, dissertation, University of Zurich, 2023

    R.J.M. Peres,Advancing Multi-Messenger Astrophysics and Dark Matter Searches with XENONnT and the Top SiPM Array of Xenoscope, dissertation, University of Zurich, 2023

  22. [41]

    CAEN,N979 - 16 Channel Fast Amplifier,https://caen.it/products/n979/, (accessed May 20, 2026). – 24 –

  23. [42]

    CAEN,VX1724 - 8 Input Channel 14 bit 100 MS/s Digitizer, https://caen.it/products/vx1724/, (accessed May 20, 2026)

  24. [43]

    Hasterok,Gas Purity Analytics, Calibration Studies, and Background Predictions towards the First Results of XENON1T, Ph.D

    C. Hasterok,Gas Purity Analytics, Calibration Studies, and Background Predictions towards the First Results of XENON1T, Ph.D. thesis, U. Heidelberg (main), 2017. 10.11588/heidok.00023693

  25. [44]

    Bakale, U

    G. Bakale, U. Sowada and W.F. Schmidt,Effect of an electric field on electron attachment to sulfur hexafluoride, nitrous oxide, and molecular oxygen in liquid argon and xenon,The Journal of Physical Chemistry80(1976) 2556

  26. [45]

    Dobi, D.S

    A. Dobi, D.S. Leonard, C. Hall, L. Kaufman, T. Langford, S. Slutsky et al.,Study of a zirconium getter for purification of xenon gas,Nucl. Instrum. Meth. A620(2010) 594 [1002.2791]

  27. [46]

    Carugno, B

    G. Carugno, B. Dainese, F. Pietropaolo and F. Ptohos,Electron lifetime detector for liquid argon, Nucl. Instrum. Meth. A292(1990) 580. [47]XENONcollaboration,The XENONnT dark matter experiment,Eur. Phys. J. C84(2024) 784 [2402.10446]. [48]DUNEcollaboration,Purity monitoring fo...

  28. [49]

    Baudis, Y

    L. Baudis, Y. Biondi, M. Galloway, F. Girard, A. Manfredini, N. McFadden et al.,Design and construction of Xenoscope — a full-scale vertical demonstrator for the DARWIN observatory,JINST 16(2021) P08052 [2105.13829]

  29. [50]

    Mount et al.,LUX-ZEPLIN (LZ) Technical Design Report,LBNL-1007256, FERMILAB-TM-2653-AE-E-PPD(2017) [1703.09144]

    B.J. Mount et al.,LUX-ZEPLIN (LZ) Technical Design Report,LBNL-1007256, FERMILAB-TM-2653-AE-E-PPD(2017) [1703.09144]

  30. [51]

    Auger et al.,The EXO-200 detector, part I: Detector design and construction,JINST7(2012) P05010 [1202.2192]

    M. Auger et al.,The EXO-200 detector, part I: Detector design and construction,JINST7(2012) P05010 [1202.2192]. [52]DarkSide-20kcollaboration,High-voltage performance testing in LAr of the PMMA cathode connection for the DarkSide-20k experiment,JINST21(2026) C04018 [2601.11837]

  31. [53]

    Saldanha et al.,Design of a high voltage delivery system for noble liquid time projection chambers, Nucl

    R. Saldanha et al.,Design of a high voltage delivery system for noble liquid time projection chambers, Nucl. Instrum. Meth. A1091(2026) 171718 [2602.23564]. – 25 –