pith. sign in

arxiv: 1507.02785 · v1 · pith:53L3ZD4Unew · submitted 2015-07-10 · ⚛️ physics.ins-det

A large open ratio, time and position sensitive detector for time of flight measurements in UHV

classification ⚛️ physics.ins-det
keywords detectordirectlyflangetimeaberrationsabsoluteaccuracyactive
0
0 comments X
read the original abstract

We report on the construction of a UHV compatible 40 mm active diameter detector based on micro channel plates and assembled directly on the feed-throughs of a DN63CF flange. It is based on the charge division technique and uses a standard two inch Si wafer as a collector. The front end electronic is placed directly on the air side of the flange allowing excellent immunity to noise and a very good timing signal with reduced ringing. The important aberrations are corrected empirically providing and absolute positioning accuracy of 500 $\mu$m while a 150 $\mu$m resolution is measured in the center.