A setup for Seebeck coefficient measurement through controlled heat pulses

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 1903.05343 v2 pith:6NHSICXC submitted 2019-03-13 physics.ins-det

classification physics.ins-det
keywords setuptemperaturewirechromelcoefficientheatlimitationmeasurement
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
abstract

A setup is designed for measuring the Seebeck coefficient (S) of materials in form of thin film, bar and wire. The main feature of this setup is control in heating and cooling cycles. In this setup heat pulse is used to generate the temperature gradient. To demonstrate the capabilities of this setup, S vs T of standard wire samples such as Au-Fe (0.07 \%), chromel, Pt and thin films of Pt and F doped SnO$_2$ are presented. The standard uncertainty of the repeatability in S measurement is found to be $\sim$ $\pm 0.056~\mu V/K$ while temperature stability is $\sim$ $\pm 10~mK$ (at 320 K), estimated for a chromel wire sample. We have tested the setup in temperature range 100 K to 320 K, while it does not have any intrinsic limitation in going down to liquid He temperatures. For temperatures above 320 K limitation is due to gluing material like varnish.