pith. sign in

arxiv: 1501.04409 · v1 · pith:EJWTJT45new · submitted 2015-01-19 · ⚛️ physics.med-ph · physics.optics

Fast Scatter Artifacts Correction for Cone-Beam CT without System Modification and Repeat Scan

classification ⚛️ physics.med-ph physics.optics
keywords scatteralgorithmartifactscbctcorrectiondatadomainfast
0
0 comments X
read the original abstract

We provide a fast and accurate scatter artifacts correction algorithm for cone beam CT (CBCT) imaging. The method starts with an estimation of coarse scatter profile for a set of CBCT images. A total-variation denoising algorithm designed specifically for Poisson signal is then applied to derive the final scatter distribution. Qualitatively and quantitatively evaluations using Monte Carlo (MC) simulations, experimental CBCT phantom data, and \emph{in vivo} human data acquired for a clinical image guided radiation therapy were performed. Results show that the proposed algorithm can significantly reduce scatter artifacts and recover the correct HU within either projection domain or image domain. Further test shows the method is robust with respect to segmentation procedure.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.