REVIEW
Properties of para-terphenyl as detector for alpha, beta and gamma radiation
Not yet reviewed by Pith; the record is open.
This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.
SPECIMEN: schema-true, not a live event
T0 review · schema-true
One-sentence machine reading of the paper's core claim.
pith:XXXXXXXX · record.json · timestamp
read the original abstract
Organic scintillators are often chosen as radiation detectors for their fast decay time and their low Z, while inorganic ones are used when high light ields are required. In this paper we show that a para-terphenyl based detector has a blend of properties of the two categories that can be optimal for energy and position measurements of low energy charged particles. On 0.1% diphenylbutadiene doped para-terphenyl samples we measure a light yield 3.5+-0.2 times larger than a typical organic scintillator (EJ-200), and a rejection power for 660 keV photons, with respect to electrons of the same energy, ranging between 3-11%, depending on the signal threshold. We also measure a light attenuation length = 4.73+-0.06 mm and we demonstrate that, with the measurements performed in this paper, a simulation based on FLUKA can properly reproduce the measured spectra.
Discussion (0). Sign in to comment.