Design and Characterization of a Novel Scintillator Array for In Vivo Monitoring During UHDR PBS Proton Therapy

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 2411.06598 v1 pith:I5CFHK5O submitted 2024-11-10 physics.med-ph

classification physics.med-ph
keywords dosearrayprotonratetherapyuhdrcorrectionimaging
verification ladder T0 review T1 audit T2 compute T3 formal

Signed reviews

No signed human review yet.

0 comments
read the original abstract

Background: Ultra-high dose rate proton therapy shows promise in tissue sparing by enhancing therapeutic ratio through the FLASH effect. In radiotherapy, accurate in vivo dosimetry is crucial for quality assurance, but remains challenging for UHDR as existing systems lack spatial and temporal resolution to verify dose and dose rate in complex anatomical regions, especially for PBS proton therapy. Purpose: To develop and evaluate a novel 3D surface dosimetry method for UHDR PBS proton therapy using high-speed imaging of a scintillator array for real-time, high-resolution surface dose monitoring. The spatial, temporal, and dosimetric components are validated via imaging of a QA phantom and comparison against TPS predictions. Methods: A deformable multi-element scintillator array was designed with 7.5mm element pitch and 0.5mm inter-element gap. Scintillation linearity with dose was evaluated with variation in response to increasing imaging and irradiation angles. WED testing evaluated beam attenuation at two energy levels. Scintillation emission was imaged at 1kHz and mesh position was monitored via 2-camera stereovision. System setup was validated using a 3D QA phantom to assess spatial accuracy and guide setup correction. Stereovision properties of array elements guided angular correction and geometric transformation. Kernel-based residual spot fitting derived cumulative dose maps compared to TPS dose profile of 5x5cm UHDR PBS delivery using 3%/2mm gamma analysis. PBS and maximum dose rate maps were calculated. Results: Setup achieved average localization error of 0.62 mm, surpassing typical 1+ mm clinical threshold. Intensity correction based on angular information yielded cumulative spot dose uncertainty of ~1% (5.428mGy). Processed dose map compared to TPS via gamma analysis showed 99.9% passing rate at 3%/2mm. WED of the array measured 1.1mm, minimizing impact on dose distribution.

Discussion (0). Continue with ORCID to comment.

Forward citations

Cited by 1 Pith paper

Reviewed papers in the Pith corpus that reference this work. Sorted by Pith novelty score. Full citation record

  1. Characterization and ex vivo application of flexible 2D scintillating coatings in ultra-high dose rate electron beams for FLASH radiotherapy

    physics.med-ph 2025-04 conditional novelty 5.0 of 10

    A flexible YAG scintillating sheet and camera system provides linear, relative real-time 2D dose maps in non-homogeneous ultra-high dose rate electron beams, including on curved ex vivo subjects, but only as relative ...