Development of the normalization method for the first large field-of-view plastic-based PET Modular scanner
Not yet reviewed by Pith; the record is open.
This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.
SPECIMEN: schema-true, not a live event
T0 review · schema-true
One-sentence machine reading of the paper's core claim.
pith:XXXXXXXX · record.json · timestamp
Signed reviews
read the original abstract
In positron emission tomography acquisition (PET), sensitivity along a line of response can vary due to crystal geometrical arrangements in the scanner and/or detector inefficiencies, leading to severe artefacts in the reconstructed image. To mitigate these effects, data must be corrected by a set of normalization coefficients applied to each line of response. The J-PET Modular scanner is a PET device made of 50 cm long plastic strips arranged axially, currently in operation at the Jagiellonian University in Krak\'ow (Poland). We have implemented a normalization method for the large field-of-view plastic-based J-PET Modular scanner using the component-based approach. We estimated the geometric normalization factors for the J-PET Modular scanner using Monte Carlo simulations. We also analysed the effects of variations in detection efficiency. A dedicated cylindrical phantom was simulated to investigate the impact of various factors on image quality. The image quality was quantified in terms of radial and axial uniformity metrics, and the standard deviation to mean intensity ratio, determined for a set of image slices. Without normalization, reconstructions of a uniform cylinder exhibit artefacts. These artefacts were satisfactorily compensated using the normalization factors. Applying geometrical corrections lowers the non-uniformity of the image expressed as a standard deviation-to-mean ratio to a range between 5.5 % to 8.5 %. Computationally, the technique is straightforward to parallelize, making it time-efficient. Preliminary estimates suggest that the method is appropriate for use with long axial field-of-view scanners, such as the total-body J-PET, currently under development at the Jagiellonian University.
Discussion (0). Continue with ORCID to comment.