Calculation of Droplet Size and Formation Time in Electrohydrodynamic Based Pulsatile Drug Delivery System

classification ⚛️ physics.med-ph
keywords dropletdrugdeliverysystemmethodsizeelectrohydrodynamicformation
0
0 comments X
read the original abstract

Electrohydrodynamic (EHD) generation, a commonly used method in BioMEMS, plays a significant role in the pulsed-release drug delivery system for a decade. In this paper, an EHD based drug delivery system is well designed, which can be used to generate a single drug droplet as small as 2.83 nL in 8.5 ms with a total device of 2x2x3 mm^3, and an external supplied voltage of 1500 V. Theoretically, we derive the expressions for the size and the formation time of a droplet generated by EHD method, while taking into account the drug supply rate, properties of liquid, gap between electrodes, nozzle size, and charged droplet neutralization. This work proves a repeatable, stable and controllable droplet generation and delivery system based on EHD method.