pith. the verified trust layer for science. sign in

arxiv: 1512.04346 · v3 · pith:QNKCAVE6new · submitted 2015-12-14 · ⚛️ physics.ins-det

Neutron spectroscopy with the Spherical Proportional Counter

classification ⚛️ physics.ins-det
keywords neutroncounterneutronsproportionalsphericalthermaldetectiondetector
0
0 comments X p. Extension
Add this Pith Number to your LaTeX paper What is a Pith Number?
\usepackage{pith}
\pithnumber{QNKCAVE6}

Prints a linked pith:QNKCAVE6 badge after your title and writes the identifier into PDF metadata. Compiles on arXiv with no extra files. Learn more

read the original abstract

A novel large volume spherical proportional counter, recently developed, is used for neutron measurements. Gas mixtures of $N_{2}$ with $C_{2}H_{6}$ and pure $N_{2}$ are studied for thermal and fast neutron detection, providing a new way for the neutron spectroscopy. The neutrons are detected via the ${}^{14}N(n, p)C^{14}$ and ${}^{14}N(n, \alpha)B^{11}$ reactions. Here we provide studies of the optimum gas mixture, the gas pressure and the most appropriate high voltage supply on the sensor of the detector in order to achieve the maximum amplification and better resolution. The detector is tested for thermal and fast neutrons detection with a ${}^{252}Cf$ and a ${}^{241}Am-{}^{9}Be$ neutron source. The atmospheric neutrons are successfully measured from thermal up to several MeV, well separated from the cosmic ray background. A comparison of the spherical proportional counter with the current available neutron counters is also given.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.