pith. sign in

arxiv: 1005.5349 · v1 · pith:QUTWXNAFnew · submitted 2010-05-28 · ⚛️ physics.ins-det

Precision timing measurement of phototube pulses using a flash analog-to-digital converter

classification ⚛️ physics.ins-det
keywords pulsesflashtimingcalorimeterforwardgluexmeasurementtime
0
0 comments X
read the original abstract

We present the timing characteristics of the flash ADC readout of the GlueX forward calorimeter, which depends on precise measurement of arrival time of pulses from FEU 84-3 photomultiplier tubes to suppress backgrounds. The tests presented were performed using two different 250 MHz prototype flash ADC devices, one with eight-bit and one with twelve-bit sampling depth. All measured time resolutions were better than 1 ns, independent of signal size, which is the design goal for the GlueX forward calorimeter. For pulses with an amplitude of 100 mV the timing resolution is 0.57 +- 0.18 ns, while for 500 mV pulses it is 0.24 +- 0.08 ns.

This paper has not been read by Pith yet.

discussion (0)