Measurement of the relative Quantum Efficiency of Hamamatsu model R5912-20MOD photomultiplier tubes at liquid argon temperature
Not yet reviewed by Pith; the record is open.
This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.
SPECIMEN: schema-true, not a live event
T0 review · schema-true
One-sentence machine reading of the paper's core claim.
pith:XXXXXXXX · record.json · timestamp
read the original abstract
The model R5912-20MOD photomultiplier tube(PMT) is made for cryogenic application by Hamamatsu. In this paper, we report on the measurement of relative quantum efficiency (QE) of this model PMT at liquid argon(LAr) temperature. Furthermore, a specially designed setup and relevant test method are introduced. The relative QE is measured in visible wavelengths with the PMT emerged in high purity nitrogen atmosphere. The results show that the change of QE at LAr temperature is within about 5% compared with room temperature around 420 nm. However, the QE increases about 10% in the shorter wavelength range and decreases significantly after 550 nm.
Forward citations
Cited by 1 Pith paper
-
Demonstration of the light collection stability of a PEN-based wavelength shifting reflector in a tonne scale liquid argon detector
A tonne-scale liquid argon detector instrumented with 4 m^2 of PEN wavelength-shifting reflector shows no light-yield deterioration over 12 days.