pith. sign in

arxiv: 1405.2089 · v1 · pith:TH356KQInew · submitted 2014-05-08 · ⚛️ physics.ins-det · hep-ex

The Performance and Long Term Stability of the D0 Run II Forward Muon Scintillation Counters

classification ⚛️ physics.ins-det hep-ex
keywords countersmuonsystemforwardscintillationamplitudeaveragecalibration
0
0 comments X
read the original abstract

The performance of the D0 experiment forward muon scintillation counters system during Run II of the Tevatron from 2001 to 2011 is described. The system consists of 4214 scintillation counters in six layers. The long term stability of the counters amplitude response determined using LED calibration system and muons produced in proton-antiproton collisions is presented. The average signal amplitude for counters of all layers has gradually decreased over ten years by 11%. The reference timing, determined using LED calibration, was stable within 0.26 ns. Average value of muon timing peak position was used for periodic D0 clock signal adjustments to compensate seasonal drift caused by temperature variations. Counters occupancy for different triggers in physics data collection runs and for minimum bias triggers are presented. The single muon yields versus time and the luminosity dependence of yields were stable for the forward muon system within 1% over 10 years.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.