pith. sign in

arxiv: 1501.04408 · v1 · pith:UQWN4LANnew · submitted 2015-01-19 · ⚛️ physics.med-ph

A patient-specific scatter artifacts correction method

classification ⚛️ physics.med-ph
keywords scattercorrectionmethodartifactartifactscbctcone-beamimaging
0
0 comments X
read the original abstract

This paper provides a fast and patient-specific scatter artifact correction method for cone-beam computed tomography (CBCT) used in image-guided interventional procedures. Due to increased irradiated volume of interest in CBCT imaging, scatter radiation has increased dramatically compared to 2D imaging, leading to a degradation of image quality. In this study, we propose a scatter artifact correction strategy using an analytical convolution-based model whose free parameters are estimated using a rough estimation of scatter profiles from the acquired cone-beam projections. It was evaluated using Monte Carlo simulations with both monochromatic and polychromatic X-ray sources. The results demonstrated that the proposed method significantly reduced the scatter-induced shading artifacts and recovered CT numbers.