Pith. sign in

REVIEW

Low noise temperature control : application to an active cavity radiometer

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv physics/0210059 v1 pith:WCLCWFDH submitted 2002-10-14 physics.ins-det

classification physics.ins-det
keywords noiseactivecavitycontrolkefflevelradiometerssensing
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
read the original abstract

We have designed low noise temperature sensing and control units with the objective of using them for the fabrication of far infrared active cavity radiometers. The sensing unit, first characterized at 300K using industrial platinium resistance thermometers, has a noise level of 25E-6 to 30E-6 Keff for a 3 hours measuring time and in a 1Hz bandwidth. Using YBCO superconducting thermometers, the noise level goes down to 2.5E-6 Keff, which is strongly limited by excess 1/f noise in the YBCO film at the superconducting transition. The sample holder used in the 90K experiments is built with an auxiliary heating resistor, which enables an easy and accurate identification of the electrothermal model, even in the closed loop operation. Based on a design previously published by NIST, we estimate from these experimental results that the overall noise limitations of radiometers could be lowered by one order of magnitude.

Discussion (0). Continue with ORCID to comment.

Pith tools