pith. sign in

arxiv: 1105.6350 · v1 · pith:WRDBM7D3new · submitted 2011-05-31 · ⚛️ physics.optics · physics.ins-det

Temperature measurement and stabilization in a birefringent whispering gallery resonator

classification ⚛️ physics.optics physics.ins-det
keywords temperaturemeasurementresonatorgallerystabilizationwhisperingactiveanisotropy
0
0 comments X
read the original abstract

Temperature measurement with nano-Kelvin resolution is demonstrated at room temperature, based on the thermal dependence of an optical crystal anisotropy in a high quality whispering gallery resonator. As the resonator's TE and TM modes frequencies have different temperature coefficients, their differential shift provides a sensitive measurement of the temperature variation, which is used for active stabilization of the temperature.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.