An affordable, wearable, fiber-free pulsed-mode diffuse speckle contrast flowmetry (PM-DSCF) sensor for noninvasive measurements of deep cerebral blood flow

T0 review · 4 major / 5 minor · reviewed 2026-08-08 · deepseek-v4-flash

Pith's one-line read Switching a wearable speckle-contrast flowmetry sensor from continuous-wave to short-pulse laser illumination extends its effective source-detector distance from 15 mm to 35 mm, corresponding to a penetration depth of about 17.5 mm, which…

desk verdict Plausible engineering step, but the 17.5-mm cerebral penetration claim is inferred, not measured, and the human data do not yet exclude scalp flow. read the letter →

arxiv 2502.08000 v1 pith:YFW7KEOO submitted 2025-02-11 q-bio.QM

classification q-bio.QM
keywords cerebralbloodflowpulsemodediffusespecklecontrastflowmetryhead-uptiltingtissuepenetrationdepthwearablesensorfiber-freenear-infrared
verification ladder T0 review T1 audit T2 compute T3 formal

The pith

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The reading

This paper aims to show that driving a diffuse speckle contrast flowmetry (DSCF) sensor with short light pulses instead of continuous illumination lets the same wearable, fiber-free probe read blood flow deeper in tissue. The authors report that the maximum usable source-detector distance grows from 15 mm in continuous-wave mode to 35 mm in pulse mode, which they interpret as an increase in penetration depth from about 7.5 mm to about 17.5 mm. That depth is enough to reach cerebral cortex through adult scalp and skull, and the paper demonstrates the device tracking cerebral blood flow changes during head-up tilting in six adults, with responses resembling those of a conventional diffuse correlation spectroscopy system. The reason this matters is that affordable, wearable, motion-tolerant monitors of deep cerebral blood flow could support continuous bedside or ambulatory monitoring, something that larger and fiber-coupled alternatives struggle to provide.

What carries the argument

The load-bearing mechanism is the pulse-mode illumination strategy: a 1 W, 808 nm laser diode is switched on for only a few milliseconds per camera exposure (duty cycle < 5%), so the instantaneous peak power is much higher than in continuous mode while the time-averaged power density stays within ANSI skin-safety limits. Higher peak power raises detected light intensity at long source-detector separations, improving signal-to-noise ratio enough that speckle contrast statistics remain usable at 35 mm. The flow readout is the standard speckle contrast $K_s = \sigma(I)/\mu(I)$ computed in a single 6400-pixel window, converted to a blood flow index $BFI = 1/K_s^2$, and compared with a baseline to give relative CBF changes.

What would settle it

A decisive test would be to measure PM-DSCF responses during head-up tilting while scalp blood flow is temporarily suppressed or removed (for example, by local pressure ischemia or a scalp tourniquet). If the 35 mm signal still shows the same tilting response, deep-brain origin is supported; if the response largely disappears, scalp blood flow is a major contributor and the claimed cerebral penetration is not established.

Watch

Extended reading notes

Core claim

Switching the DSCF laser from continuous-wave to short-pulse operation, with a duty cycle below 5% and average power density at or below 0.28 W/cm², increases the maximum effective source-detector distance from 15 mm to 35 mm while keeping skin exposure within the ANSI Z136.1 limit. The authors verify on tissue-simulating phantoms that pulse-mode measurements at 30-40 mm source-detector separation track temperature-driven flow changes, whereas continuous-wave measurements at those distances do not, and they show that in six healthy adults a 35 mm source-detector PM-DSCF probe detects head-up-tilting cerebral blood flow responses that correlate significantly with concurrent DCS measurements (R² = 0.75). They conclude that pulse-mode DSCF reaches about 17.5 mm penetration depth, sufficient for adult human cerebral blood flow monitoring.

Load-bearing premise

The claim depends on the assumption that tightening the headband pushes scalp blood away from the probe, so the light detected at 35 mm comes from brain tissue rather than scalp; the paper cites earlier work for this but does not independently verify it in these measurements.

Editorial extensions

If this is right

  • PM-DSCF at a 35 mm source-detector distance can track cerebral blood flow changes in adult humans, a capability the same wearable probe did not have in continuous-wave mode.
  • Pulse-mode operation at 30-40 mm source-detector separations detects dynamic flow changes in tissue-simulating phantoms, while continuous-wave operation at those distances does not.
  • The 70° head-up-tilting responses measured by PM-DSCF and a conventional DCS system are significantly correlated (R² = 0.75), with PM-DSCF showing at least as large responses, consistent with deeper penetration.
  • Optimized camera exposure time is required for deep measurements: at 35 mm, exposures of 10 ms or longer give reliable flow signals, while 1-5 ms exposures lose sensitivity.
  • The skin-safety budget is maintained in pulse mode: average power density stays below the ANSI limit, and no scalp temperature increases or subject complaints were reported during the study.

Reading between the lines

Editorial extensions of the paper, not claims the author makes directly.

  • Beyond the paper: if the penetration gain is real, the same headband could carry two DSCF channels at short and long source-detector distances, letting one separate scalp blood flow from cerebral blood flow without changing hardware.
  • Beyond the paper: the pulse-mode gain should be directly tested against the scalp-removal assumption by comparing 35 mm signals during head-up tilting with and without local scalp ischemia; the paper does not include such a control.
  • Beyond the paper: a faster camera or higher duty cycle could push the 1.65 Hz sampling rate toward cardiac-frequency measurements, but this is not demonstrated and would need a new safety evaluation.
  • Beyond the paper: because the phantom absorption coefficient used here (0.03 cm⁻¹) is lower than typical adult tissue, real-tissue performance may be somewhat worse; a layered phantom or Monte Carlo simulation with realistic scalp and skull optics would bound the actual depth.
Share X Bluesky LinkedIn Reddit HN

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, and a circularity audit.

Referee Report

4 major / 5 minor

Summary. The manuscript reports a pulsed-mode diffuse speckle contrast flowmetry (PM-DSCF) system that uses an 808 nm laser diode operated with a duty cycle below 5% and a compact NanEye camera to measure deep cerebral blood flow in adult humans. The authors compare PM-DSCF with their earlier CW-DSCF on tissue-simulating phantoms, reporting that the maximum effective source-detector distance increases from 15 mm to 35 mm, which they translate into a penetration depth of about 17.5 mm. They then present concurrent PM-DSCF and DCS measurements during 70° head-up tilting in six healthy adults, showing a group-average rCBF increase of about 45% with a linear correlation R² = 0.75 between the two techniques. The central claim is that switching from CW to pulsed illumination with optimized exposure time enables noninvasive CBF measurements in adult human brains with a wearable, fiber-free, low-cost sensor.

Significance. If the claimed penetration depth and cerebral specificity are substantiated, this work would represent a practical advance: it combines a low-cost, wearable, fiber-free design with a larger source-detector distance than typical DCS systems, potentially enabling continuous CBF monitoring in adults outside specialized labs. The manuscript also provides useful phantom data comparing CW and pulsed operation, including SNR as a function of source-detector distance and exposure time, and it documents dynamic flow detection at 35-40 mm in a tissue-simulating phantom. The authors are appropriately explicit about some limitations, such as the lower-than-realistic phantom absorption coefficient and the conventional speckle contrast analysis without full noise characterization. However, the significance is currently tempered by the absence of a direct validation that the measured signal originates from cerebral tissue rather than scalp, which is the load-bearing point for the headline claim of adult CBF measurement.

major comments (4)
  1. [Section 2.2 and Fig. 3] The phantom experiments that establish the 35 mm effective S-D distance use an absorption coefficient of μa = 0.03 cm⁻¹, which the authors themselves note is lower than typical realistic tissue absorption. Since the SNR at large S-D distances in real tissue will be degraded by higher absorption, the Fig. 3 results do not by themselves demonstrate that 35 mm S-D is viable on an adult head. Please provide either measurements on a phantom with tissue-realistic optical properties (e.g., μa ≈ 0.1-0.2 cm⁻¹ at 808 nm) or a quantitative model-based estimate of the expected SNR at 35 mm for adult scalp/skull/brain layers.
  2. [Section 4, Abstract] The penetration depth of approximately 17.5 mm is inferred as half of the 35 mm S-D distance, but this linear relation is not established for the layered adult head. Diffuse optical sensitivity profiles for realistic head models peak near the surface and do not simply extend to S-D/2. Please support the penetration-depth claim with a layered Monte Carlo simulation, a layered phantom measurement, or an explicit sensitivity-profile calculation; otherwise, the abstract's statement about 'deep cerebral blood flow' is overstated.
  3. [Section 2.3 and Fig. 6] The adult validation compares PM-DSCF on the right forehead at 35 mm S-D with DCS on the left forehead at 25 mm S-D. Because head-up tilting is a systemic autonomic challenge, a correlation between two non-co-localized measurements does not establish that PM-DSCF is specifically sensing cerebral blood flow. The R² = 0.75 could be driven by simultaneous changes in scalp blood flow. Please provide co-localized PM-DSCF and DCS measurements, or an independent control such as superficial scalp occlusion or a layered head phantom, to separate cerebral from extracerebral contributions.
  4. [Section 2.3] The only argument excluding scalp blood flow is the statement that tightening the headband 'displaces scalp blood away from the probe,' supported by a citation to reference 49. No pressure calibration, no scalp-flow measurement, and no quantification of the residual extracerebral contribution are provided. Given that the observed HUT response is about 45% and that scalp blood flow can respond to posture and pressure, this assumption needs direct verification before the conclusion 'successful CBF measurements in adult humans' can be accepted.
minor comments (5)
  1. [Section 2.1.3] The choice of the 6,400-pixel single window over the sliding-window averaging approach is justified only by 'data not shown'; please include the supporting comparison or provide a reference to a prior publication.
  2. [Fig. 3 caption] Please define explicitly how SNR is computed from the intensity and dark-count data, and state the threshold criterion (e.g., the 6 dB and 12 dB dashed lines) in the main text as well as the caption.
  3. [Section 2.1.2] The synchronization between the pulsed laser and the NanEye camera is described qualitatively; please specify the timing jitter and how 'noticeable noises caused by misalignment' were detected and excluded, since this affects the validity of the speckle contrast estimates.
  4. [Throughout] Several typographical errors remain, including 'custimized' in Fig. 1 caption, 'temprature' and 'manuplate' in Fig. 2 caption, 'continously' in Fig. 2 caption, and 'HTU' instead of 'HUT' in the Discussion; a careful proofreading pass is needed.
  5. [Section 4, pulse broadening discussion] The discussion of pulse broadening and coherence length is helpful, but it would benefit from citing specific quantitative coherence-length measurements from reference 65 rather than only stating that the effects are 'not major concerns.'

Circularity Check

2 steps flagged · score 6.0 of 10

Core pulse-mode SNR improvement is empirically supported, but the headline 17.5-mm penetration depth is just S-D/2 by the paper's own definition, and the human HUT validation uses a processing window chosen for stronger HUT/DCS agreement.

  1. self definitional [Section 4 (Discussion and Conclusions), final paragraph; cf. Section 1 definition of penetration depth as one-half of the S-D distance]
    "Correspondingly, the maximum penetration depth of PM-DSCF reaches approximately 17.5 mm (half of 35 mm)."

    The paper elsewhere defines maximal penetration depth as one-half of the S-D distance ('maximal penetration depth of ~12.5 mm (i.e., one-half of the S-D distance)'), so the statement that a 35-mm S-D distance gives ~17.5-mm penetration is not an independent measurement or prediction. It is the same quantity as the chosen S-D distance under the paper's own definition, divided by two. The phantom SNR data can show that usable signals exist at 35 mm, but the '17.5 mm penetration depth' conclusion is a restatement of the input geometry, not a derived result.

  2. fitted input called prediction [Section 2.1.3 (Data Processing)]
    "Based on our comparison results from human studies (data not shown), the single-window approach exhibited stronger rCBF responses to HUT compared to the sliding-window averaging approach. Furthermore, rCBF responses measured using the single window approach closely aligned with DCS measurements. Consequently, this study adopted a single window of 6,400 pixels for Ks calculation."

    The single 80x80-pixel window was adopted because it produced larger HUT responses and closer DCS agreement in the same human data. Those same HUT responses and the R2 = 0.75 DCS correlation (Fig. 6) are then presented as validation that PM-DSCF measures adult CBF. Because the processing choice was selected on the target outcomes, the reported response amplitude and agreement are in part determined by the selection criterion, making the validation partly circular. The comparison underlying the selection is 'data not shown', so the effect cannot be independently checked.

full rationale

No circularity is found in the core instrument evaluation: the CW-vs-PM comparison on phantoms (Figs. 3-4, Table 1) is a direct empirical SNR/sensitivity test against a temperature-induced flow change, and BFI/rCBF follow standard speckle-contrast definitions rather than being fitted to the outcome. The increase in usable S-D distance from 15 to 35 mm is an experimental observation, not a consequence of the claims. However, two load-bearing presentation steps are circular or self-definitional. First, the advertised '~17.5 mm penetration depth' is obtained by applying the paper's own one-half-S-D rule to the chosen 35-mm distance, so it carries no independent evidential content. Second, the human HUT validation uses a data-processing window that was explicitly chosen (in the same dataset) for stronger HUT response and closer DCS correlation, so the subsequent Fig. 6 agreement is partly forced by that selection. The scalp-exclusion argument and the non-co-localized contralateral DCS comparison are validity concerns rather than circularity themselves, but they compound the fragility. Overall, these issues do not erase the phantom-based engineering advance, but they make the 'deep CBF in adults' claim partially circular in its current presentation. Self-citations to the authors' prior CW-DSCF work are used as baseline context, not as proof of the new PM-mode performance, and therefore do not independently raise the circularity score.

Assumptions & free parameters 4 free parameters · 4 assumptions · 0 invented entities

The central claim relies on a small number of free parameters (SNR threshold, exposure time, ROI size, duty cycle) that were tuned in this study, plus four domain assumptions about speckle contrast theory, the half-distance depth rule, scalp blood displacement by headband pressure, and phantom-tissue equivalence. No new physical entities are introduced.

free parameters (4)
  • Minimum SNR threshold = 6 dB
    Used to define 'valid' S-D distances in phantom experiments; choosing a different threshold would change the claimed maximum distance.
  • Camera exposure time = 10 ms
    Selected for human measurements based on a single case study (Subject #4); sensitivity results depend on this choice.
  • ROI pixel count = 6400 pixels
    The 80x80 ROI was chosen after unpublished comparisons showing stronger HUT responses than sliding-window averaging.
  • Laser duty cycle = <5%
    Chosen to keep average power density at 0.28 W/cm2 under the ANSI limit while maximizing peak power; the specific value influences SNR.
assumptions (4)
  • domain assumption Speckle contrast Ks = sigma/mu, with blood flow index BFI = 1/Ks^2, monotonically reflects blood flow
    Standard laser speckle contrast theory used throughout; the paper defines BFI this way in Section 2.1.3.
  • domain assumption Maximum tissue penetration depth equals half the source-detector distance
    Conventional approximation in diffuse optics, used to convert 35 mm S-D to 17.5 mm depth in the abstract and discussion.
  • domain assumption Tightly applied headband displaces scalp blood such that the measured signal is dominated by cerebral blood flow
    Stated in Section 2.3 and based on a prior study (ref 49); not directly verified in this dataset.
  • domain assumption Liquid phantom with mu_a=0.03 cm^-1 and mu_s'=9 cm^-1 is representative enough to predict in vivo performance
    The paper itself notes the absorption is lower than typical tissue, which may make the phantom more permissive than real heads; stated in Section 2.2.

how reviews work

0 comments
Cite this review

Pith. "Pith review of An affordable, wearable, fiber-free pulsed-mode diffuse speckle contrast flowmetry (PM-DSCF) sensor for noninvasive measurements of deep cerebral blood flow." pith.science (2026). https://pith.science/paper/YFW7KEOO

@misc{pith2026250208000,
  author       = {Pith},
  title        = {Pith review of: An affordable, wearable, fiber-free pulsed-mode diffuse speckle contrast flowmetry (PM-DSCF) sensor for noninvasive measurements of deep cerebral blood flow},
  year         = {2026},
  howpublished = {\url{https://pith.science/paper/YFW7KEOO}},
  note         = {Machine review of arXiv:2502.08000}
}
read the original abstract

Significance: Measuring cerebral blood flow (CBF) is crucial for diagnosing various cerebral diseases. An affordable, wearable, and fiber-free continuous-wave speckle contrast flowmetry (CW-DSCF) technique has been developed for continuous monitoring of CBF variations. However, its application in adult humans is limited by shallow tissue penetration. Aim: To develop an innovative pulse-mode DSCF (PM-DSCF) system for continuous monitoring of CBF variations in adult humans. Approach: The PM-DSCF utilizes an 808 nm laser diode and a small NanEye camera to capture diffuse laser speckle fluctuations caused by red blood cell movement in the brain (i.e., CBF). Operating in short-pulse mode (duty cycle < 5%), the system maximizes peak pulse light power for deeper tissue penetration, while ensuring that the average power density remains within ANSI safety standards for skin exposure. The PM-DSCF was evaluated on tissue-simulating phantoms and in adult humans. Results: The maximum effective source-detector distance increased from 15 mm (CW-DSCF) to 35 mm (PM-DSCF). The PM-DSCF successfully detected CBF variations in adult brains during head-up-tilting experiments, consistent with physiological expectations. Conclusions: Switching from CW mode to PM mode significantly increases the maximum tissue penetration depth from ~7.5 mm (CW-DSCF) to ~17.5 mm (PM-DSCF), enabling successful CBF measurements in adult humans.

Figures

Figures reproduced from arXiv: 2502.08000 by the authors.

Figure 1
Figure 1. An affordable, wearable, fiber-free PM-DSCF system for deep tissue blood flow measurements. (a) A headband of PM-DSCF placed on the subject’s forehead. The PM-DSCF device included a user interface ①, an Arduino controller ②, a custimized laser diode driver ③, a laser diode ④, a camera board (NanEye USB 2.2, Awaiba) ⑤, and a NanEye camera ⑥. (b) A schematic diagram of PM-DSCF device. (c) Synchronization of the pulsed… view at source ↗
Figure 2
Figure 2. Experimental setup for evaluating the performance of PM-DSCF in tissue-simulating phantoms and human adults. (a) A PM-DSCF probe was placed on the surface of Intralipid liquid phantom with known optical properties [PITH_FULL_IMAGE:figures/full_fig_p010_2.png] view at source ↗
Figure 3
Figure 3. Comparison of CW-DSCF and PM-DSCF measurement sensitivities and SNRs on tissue-simulating phantoms with varied S-D distances and camera exposure times. (a) Average light intensity (digital counts) and dark-noise counts detected by the CW-DSCF. (b) Average light intensity (digital counts) and dark-noise counts detected by the PM-DSCF. (c) Comparison of measurement SNRs achieved by the CW-DSCF and PM-DSCF. The two das… view at source ↗
Figures from the paper (7 more)
Figure 4
Figure 4. Figure 4: CW-DSCF and PM-DSCF measurements of dynamic flow changes (rBFI) in tissue-simulating phantoms, induced by changing phantom temperature. (a)-(c) CW-DSCF measurements of rBFIs (means ± standard deviations) with S-D distances of 30, 35, and 40 mm, respectively. (d)-(f) PM…
Figure 5
Figure 5. Figure 5: shows the concurrent DCS and PM-DSCF measurement results of rCBF changes in left and right hemispheres of a illustrative subject during 70° HUT, using the S-D distances of 25, 30, and 35 mm, respectively. The camera exposure time was set to 10 ms for PM-DSCF measuremen…
Figure 6
Figure 6. Figure 6: Group average results from concurrent DCS and PM-DSCF measurements of rCBF changes in left and right hemispheres of 6 human adults during 70° HUT. (a) Time-course changes of left-rCBF (by DCS with the S-D distance of 25 mm) and right-rCBF (by PM-DSCF with the camera ex…
Figure 1
Figure 1. Figure 1: An affordable, wearable, fiber-free PM-DSCF system for deep tissue blood flow measurements. (a) A headband of PM-DSCF placed on the subject’s forehead. The PM-DSCF device included a user interface ①, an Arduino controller ②, a custimized laser diode driver ③, a laser d…
Figure 2
Figure 2. Figure 2: Experimental setup for evaluating the performance of PM-DSCF in tissue-simulating phantoms and human adults. (a) A PM-DSCF probe was placed on the surface of Intralipid liquid phantom with known optical properties to test and compare the measurement sensitivities and S…
Figure 3
Figure 3. Figure 3: Comparison of CW-DSCF and PM-DSCF measurement sensitivities and SNRs on tissue￾simulating phantoms with varied S-D distances and camera exposure times. (a) Average light intensity (digital counts) and dark-noise counts detected by the CW-DSCF. (b) Average light intensi…
Figure 5
Figure 5. Figure 5: Concurrent DCS and PM-DSCF measurements of rCBF changes in left and right hemispheres of a human adult (Subject #4) during 70° HUT, with S-D distances of (a) 25 mm, (b) 30 mm, and (c) 35 mm, respectively [PITH_FULL_IMAGE:figures/full_fig_p029_5.png]

Discussion (0). Continue with ORCID to comment.

Reference graph

Works this paper leans on

66 extracted references · 57 canonical work pages

  1. [1]

    An integrated approach to measuring tumor oxygen status using human melanoma xenografts as a model,

    C. Menon et al., "An integrated approach to measuring tumor oxygen status using human melanoma xenografts as a model," Cancer Res. 63 (21), 7232-7240 (2003)

  2. [2]

    Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation,

    D. A. Boas and A. G. Yodh, "Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation," JOSA A 14 (1), 192 -215 (1997). https://doi.org/10.1364/JOSAA.14.000192

  3. [3]

    Diffuse optical tomography of cerebral blood flow, oxygenation, and metabolism in rat during focal ischemia,

    J. P. Culver et al., "Diffuse optical tomography of cerebral blood flow, oxygenation, and metabolism in rat during focal ischemia," J Cereb Blood Flow Metab. 23 (8), 911-924 (2003). https://doi.org/10.1097/01.WCB.0000076703.71231.BB

  4. [4]

    L. G. Morris and J. P. Shah, Technical variations and refinements in head and neck surgery, JP Medical Ltd (2014)

  5. [5]

    Automatic analysis of skull thickness, scalp -to-cortex distance and association with age and sex in cognitively normal elderly,

    J. Zhang et al., "Automatic analysis of skull thickness, scalp -to-cortex distance and association with age and sex in cognitively normal elderly," bioRxiv 16 (2), 653 -656 (2023). https://doi.org/10.1101/2023.01.19.524484

  6. [6]

    A wearable fiber -free optical sensor for continuous monitoring of cerebral blood flow in freely behaving mice,

    X. Liu et al., "A wearable fiber -free optical sensor for continuous monitoring of cerebral blood flow in freely behaving mice," IEEE Trans Biomed Eng. 70 (6), 1838-1848 (2023). https://doi.org/10.1109/TBME.2022.3229513

  7. [7]

    Diffuse correlation spectroscopy for measurement of cerebral blood flow: Future prospects,

    E. M. Buckley et al., "Diffuse correlation spectroscopy for measurement of cerebral blood flow: Future prospects," Neurophotonics 1 (1), (2014). https://doi.org/10.1117/1.NPh.1.1.011009

  8. [8]

    Novel experimental model for repeated forebrain ischemia -reperfusion,

    L. Chen et al., "Novel experimental model for repeated forebrain ischemia -reperfusion," Journal of experimental stroke and translational medicine 5 (1), 1 -10 (2012). https://doi.org/10.4172/1939-067X.1000136

Show all 66 references
  1. [9]

    Noninvasive optical evaluation of spontaneous low frequency oscillations in cerebral hemodynamics,

    R. Cheng et al., "Noninvasive optical evaluation of spontaneous low frequency oscillations in cerebral hemodynamics," Neuroimage 62 (3), 1445 -1454 (2012). https://doi.org/10.1016/j.neuroimage.2012.05.069

  2. [10]

    Near -infrared diffuse optical monitoring of cerebral blood flow and oxygenation for the prediction of vasovagal syncope,

    R. Cheng et al., "Near -infrared diffuse optical monitoring of cerebral blood flow and oxygenation for the prediction of vasovagal syncope," J Biomed Opt. 19 (1), (2014). https://doi.org/10.1117/1.JBO.19.1.017001

  3. [11]

    Diffuse correlation spectroscopy for non -invasive, micro- vascular cerebral blood flow measurement,

    T. Durduran and A. G. Yodh, "Diffuse correlation spectroscopy for non -invasive, micro- vascular cerebral blood flow measurement," Neuroimage 85 Pt 1 (0 1), 51 -63 (2014). https://doi.org/10.1016/j.neuroimage.2013.06.017

  4. [12]

    Dynamic cerebral autoregulation measured by diffuse correlation spectroscopy,

    C. G. Favilla et al., "Dynamic cerebral autoregulation measured by diffuse correlation spectroscopy," J Cereb Blood Flow Metab 43 (8), 1317 -1327 (2023). https://doi.org/10.1177/0271678X231153728

  5. [13]

    Noninvasive measurement of cerebral blood flow and blood oxygenation using near -infrared and diffuse correlation spectroscopies in critically brain -injured adults,

    M. N. Kim et al., "Noninvasive measurement of cerebral blood flow and blood oxygenation using near -infrared and diffuse correlation spectroscopies in critically brain -injured adults," Neurocrit Care 12 (2), 173-180 (2010). https://doi.org/10.1007/s12028-009-9305- x

  6. [14]

    Diffuse optical monitoring of repeated cerebral ischemia in mice,

    Y. Shang et al., "Diffuse optical monitoring of repeated cerebral ischemia in mice," Opt Express. 19 (21), 20301-20315 (2011). https://doi.org/10.1364/OE.19.020301

  7. [15]

    Cerebral monitoring during carotid endarterectomy using near -infrared diffuse optical spectroscopies and electroencephalogram,

    Y. Shang et al., "Cerebral monitoring during carotid endarterectomy using near -infrared diffuse optical spectroscopies and electroencephalogram," Phys Med Biol. 56 (10), 3015- 3032 (2011). https://doi.org/10.1088/0031-9155/56/10/008. 24

  8. [16]

    Highly parallel, interferometric diffusing wave spectroscopy for monitoring cerebral blood flow dynamics,

    W. Zhou et al., "Highly parallel, interferometric diffusing wave spectroscopy for monitoring cerebral blood flow dynamics," Optica 5 (5), 518 -527 (2018). https://doi.org/10.1364/OPTICA.5.000518

  9. [17]

    Y. X. Huang et al., "Interferometric speckle visibility spectroscopy (i SVS) for measuring decorrelation time and dynamics of moving samples with enhanced signal -to-noise ratio and relaxed reference requirements," Opt Express 31 (19), 31253 -31266 (2023). https://doi.org/10.1...

  10. [18]

    Portable, high speed blood flow measurements enabled by long wavelength, interferometric diffuse correlation spectroscopy (LW-iDCS),

    M. B. Robinson et al., "Portable, high speed blood flow measurements enabled by long wavelength, interferometric diffuse correlation spectroscopy (LW-iDCS)," Sci Rep. 13 (1), 8803 (2023). https://doi.org/10.1038/s41598-023-36074-8

  11. [19]

    Fast time -domain diffuse correlation spectroscopy with superconducting nanowire single-photon detector: System validation and in vivo results,

    V. Parfentyeva et al., "Fast time -domain diffuse correlation spectroscopy with superconducting nanowire single-photon detector: System validation and in vivo results," Sci Rep. 13 (1), 11982 (2023). https://doi.org/10.1038/s41598-023-39281-5

  12. [20]

    Time -domain diffuse correlation spectroscopy,

    J. Sutin et al., "Time -domain diffuse correlation spectroscopy," Optica 3 (9), 1006-1013 (2016). https://doi.org/10.1364/OPTICA.3.001006

  13. [21]

    Analytical models for time-domain diffuse correlation spectroscopy for multi- layer and heterogeneous turbid media,

    J. Li et al., "Analytical models for time-domain diffuse correlation spectroscopy for multi- layer and heterogeneous turbid media," Biomed Opt Express 8 (12), 5518-5532 (2017). https://doi.org/10.1364/BOE.8.005518

  14. [22]

    Portable optical tissue flow oximeter based on diffuse correlation spectroscopy,

    Y. Shang et al., "Portable optical tissue flow oximeter based on diffuse correlation spectroscopy," Opt Lett. 34 (22), 3556 -3558 (2009). https://doi.org/10.1364/OL.34.003556

  15. [23]

    Noninvasive optical quantification of absolute blood flow, blood oxygenation, and oxygen consumption rate in exercising skeletal muscle,

    K. Gurley, Y. Shang and G. Yu, "Noninvasive optical quantification of absolute blood flow, blood oxygenation, and oxygen consumption rate in exercising skeletal muscle," J Biomed Opt. 17 (7), 075010 (2012). https://doi.org/10.1117/1.JBO.17.7.075010

  16. [24]

    Extraction of diffuse correlation spectroscopy flow index by integration of nth-order linear model with monte carlo simulation,

    Y. Shang et al., "Extraction of diffuse correlation spectroscopy flow index by integration of nth-order linear model with monte carlo simulation," Appl Phys Lett. 104 (19), 193703 (2014). https://doi.org/10.1063/1.4876216

  17. [25]

    A nth -order linear algorithm for extracting diffuse correlation spectroscopy blood flow indices in heterogeneous tissues,

    Y. Shang and G. Yu, "A nth -order linear algorithm for extracting diffuse correlation spectroscopy blood flow indices in heterogeneous tissues," Appl Phys Lett. 105 (13), 133702 (2014). https://doi.org/10.1063/1.4896992

  18. [26]

    Wearable fiber -free optical sensor for continuous monitoring of neonatal cerebral blood flow and oxygenation,

    X. Liu et al., "Wearable fiber -free optical sensor for continuous monitoring of neonatal cerebral blood flow and oxygenation," Pediatric Research , 1 -8 (2024). https://doi.org/10.1101/2023.09.21.23295914

  19. [27]

    Deep tissue flowmetry based on diffuse speckle contrast analysis,

    R. Bi, J. Dong and K. Lee, "Deep tissue flowmetry based on diffuse speckle contrast analysis," Opt Lett 38 (9), 1401-1403 (2013). https://doi.org/10.1364/OL.38.001401

  20. [28]

    Compact, multi-exposure speckle contrast optical spectroscopy (scos) device for measuring deep tissue blood flow,

    T. Dragojević et al., "Compact, multi-exposure speckle contrast optical spectroscopy (scos) device for measuring deep tissue blood flow," Biomed Opt Express. 9 (1), 322-334 (2018). https://doi.org/10.1364/BOE.9.000322

  21. [29]

    Validation of the openwater wearable optical system: Cerebral hemodynamic monitoring during a breath -hold maneuver,

    C. G. Favilla et al., "Validation of the openwater wearable optical system: Cerebral hemodynamic monitoring during a breath -hold maneuver," Neurophotonics 11 (1), 015008-015008 (2024). https://doi.org/10.1117/1.NPh.11.1.015008

  22. [30]

    A wearable fiberless optical sensor for continuous monitoring of cerebral blood flow in mice,

    C. Huang et al., "A wearable fiberless optical sensor for continuous monitoring of cerebral blood flow in mice," IEEE J Sel Top Quantum Electron 25 (1), (2019). https://doi.org/10.1109/JSTQE.2018.2854597. 25

  23. [31]

    Low -cost compact diffuse speckle contrast flowmeter using small laser diode and bare charge -coupled-device,

    C. Huang et al., "Low -cost compact diffuse speckle contrast flowmeter using small laser diode and bare charge -coupled-device," J Biomed Opt. 21 (8), 080501 -080501 (2016). https://doi.org/10.1117/1.JBO.21.8.080501

  24. [32]

    Measuring human cerebral blood flow and brain function with fiber -based speckle contrast optical spectroscopy system,

    B. Kim et al., "Measuring human cerebral blood flow and brain function with fiber -based speckle contrast optical spectroscopy system," Commun Biol. 6 (1), 844 (2023). https://doi.org/10.1038/s42003-023-05211-4

  25. [33]

    Multi -mode fiber-based speckle contrast optical spectroscopy: Analysis of speckle statistics,

    C. P. Lin et al., "Multi -mode fiber-based speckle contrast optical spectroscopy: Analysis of speckle statistics," Opt Lett 48 (6), 1427 -1430 (2023). https://doi.org/10.1364/OL.478956

  26. [34]

    Simultaneous measurements of tissue blood flow and oxygenation using a wearable fiber -free optical sensor.,

    X. Liu et al., "Simultaneous measurements of tissue blood flow and oxygenation using a wearable fiber -free optical sensor.," J Biomed Opt. 26 (1), 012705 (2021). https://doi.org/10.1117/1.JBO.26.1.012705

  27. [36]

    Speckle contrast optical spectroscopy, a non -invasive, diffuse optical method for measuring microvascular blood flow in tissue,

    C. P. Valdes et al., "Speckle contrast optical spectroscopy, a non -invasive, diffuse optical method for measuring microvascular blood flow in tissue," Biomed Opt Express 5 (8), 2769-2784 (2014). https://doi.org/10.1364/BOE.5.002769

  28. [37]

    Diffusing wave spectroscopy: A unified treatment on temporal sampling and speckle ensemble methods,

    J. Xu, A. K. Jahromi and C. Yang, "Diffusing wave spectroscopy: A unified treatment on temporal sampling and speckle ensemble methods," APL Photonics 6 (1), (2021). https://doi.org/10.1063/5.0034576

  29. [38]

    Compact low -cost fiberless diffuse speckle contrast flow-oximeter,

    G. Yu, C. Huang and T. Hastings, "Compact low -cost fiberless diffuse speckle contrast flow-oximeter," US Patent# 10/842,422, (2020)

  30. [39]

    Laser speckle contrast imaging: Theoretical and practical limitations,

    D. Briers et al., "Laser speckle contrast imaging: Theoretical and practical limitations," J Biomed Opt. 18 (6), 066018-066018 (2013). https://doi.org/10.1117/1.JBO.18.6.066018

  31. [40]

    Measurement sensitivity refers to the ability of DSCF device to detect small changes in CBF. We employed a pulse -mode strategy in the DSCF system that utilize d high-intensity illumination for a short pulse duration while maintaining the average power density below the ANSI s...

  32. [41]

    Determination of optimal exposure time for imaging of blood flow changes with laser speckle contrast imaging,

    S. Yuan et al., "Determination of optimal exposure time for imaging of blood flow changes with laser speckle contrast imaging," Appl Opt. 44 (10), 1823 -1830 (2005). https://doi.org/10.1364/ao.44.001823

  33. [42]

    Speckle contrast diffuse correlation tomography of complex turbid medium flow,

    C. Huang et al., "Speckle contrast diffuse correlation tomography of complex turbid medium flow," Med Phys. 42 (7), 4000 -4006 (2015). https://doi.org/doi:10.1118/1.4922206

  34. [43]

    Noncontact 3-d speckle contrast diffuse correlation tomography of tissue blood flow distribution,

    C. Huang et al., "Noncontact 3-d speckle contrast diffuse correlation tomography of tissue blood flow distribution," IEEE Trans Med Imaging 36 (10), 2068 -2076 (2017). https://doi.org/10.1109/TMI.2017.2708661

  35. [44]

    Diffuse speckle contrast analysis (dsca) for deep tissue blood flow monitoring,

    K. Lee, "Diffuse speckle contrast analysis (dsca) for deep tissue blood flow monitoring," Advanced Biomedical Engineering 9, 21-30 (2020). https://doi.org/10.14326/abe.9.21

  36. [45]

    Simultaneous blood flow and blood oxygenation measurements using a combination of diffuse speckle contrast analysis and near-infrared spectroscopy,

    M. Seong et al., "Simultaneous blood flow and blood oxygenation measurements using a combination of diffuse speckle contrast analysis and near-infrared spectroscopy," J Biomed Opt. 21 (2), 027001-027001 (2016). https://doi.org/10.1117/1.JBO.21.2.027001

  37. [46]

    Fast pulsatile blood flow measurement in deep tissue through a multimode detection fiber,

    R. Bi et al., "Fast pulsatile blood flow measurement in deep tissue through a multimode detection fiber," J Biomed Opt. 25 (5), 055003 -055003 (2020). https://doi.org/10.1117/1.JBO.25.5.055003

  38. [47]

    Pressure stimulus study on acupuncture points with multi- channel multimode -fiber diffuse speckle contrast analysis (mmf -dsca),

    J. Y. Park, G. Choi and K. Lee, "Pressure stimulus study on acupuncture points with multi- channel multimode -fiber diffuse speckle contrast analysis (mmf -dsca)," Biomed Opt Express 14 (11), 5602-5614 (2023). https://doi.org/10.1364/BOE.502447. 26

  39. [48]

    Noncontact diffuse optical assessment of blood flow changes in head and neck free tissue transfer flaps,

    C. Huang et al., "Noncontact diffuse optical assessment of blood flow changes in head and neck free tissue transfer flaps," J Biomed Opt. 20 (7), 075008 -075008 (2015). https://doi.org/10.1117/1.JBO.20.7.075008

  40. [49]

    Three -dimensional flow contrast imaging of deep tissue using noncontact diffuse correlation tomography,

    Y. Lin et al., "Three -dimensional flow contrast imaging of deep tissue using noncontact diffuse correlation tomography," Appl Phys Lett. 104 (12), (2014). https://doi.org/10.1063/1.4869469

  41. [50]

    Influence of probe pressure on the pulsatile diffuse correlation spectroscopy blood flow signal on the forearm and forehead regions,

    D. Wang et al., "Influence of probe pressure on the pulsatile diffuse correlation spectroscopy blood flow signal on the forearm and forehead regions," Neurophotonics 6 (3), 035013 (2019). https://doi.org/10.1117/1.NPh.6.3.035013

  42. [51]

    Beneath the surface: Revealing deep-tissue blood flow in human subjects with massively parallelized diffuse correlation spectroscopy,

    L. Kreiss et al., "Beneath the surface: Revealing deep-tissue blood flow in human subjects with massively parallelized diffuse correlation spectroscopy," arXiv preprint , arXiv:2403.03968 (2024). https://doi.org/10.48550/arXiv.2403.03968

  43. [52]

    Quantification of cerebral blood flow in adults by contrast-enhanced near- infrared spectroscopy: Validation against mri,

    D. Milej et al., "Quantification of cerebral blood flow in adults by contrast-enhanced near- infrared spectroscopy: Validation against mri," J Cereb Blood Flow Metab 40 (8), 1672- 1684 (2020). https://doi.org/10.1177/0271678X19872564

  44. [53]

    I the statistics of speckle patterns,

    J. C. Dainty, "I the statistics of speckle patterns," Progress in Optics 14, 1 -46 (1977). https://doi.org/10.1016/S0079-6638(08)70249-X

  45. [54]

    J. P. Culver et al., "Three -dimensional diffuse optical tomography in the parallel plane transmission geometry: Evaluation of a hybrid frequency domain/continuous wave clinical system for breast imaging," Med Phys 30 (2), 235 -247 (2003). https://doi.org/10.1118/1.1534109

  46. [55]

    Hemodynamic evoked response of the sensorimotor cortex measured noninvasively with near -infrared optical imaging,

    M. A. Franceschini et al., "Hemodynamic evoked response of the sensorimotor cortex measured noninvasively with near -infrared optical imaging," Psychophysiology 40 (4), 548-560 (2003). https://doi.org/10.1111/1469-8986.00057

  47. [56]

    High -density diffuse optical tomography for imaging human brain function,

    M. D. Wheelock, J. P. Culver and A. T. Eggebrecht, "High -density diffuse optical tomography for imaging human brain function," Rev Sci Instrum 90 (5), 051101 (2019). https://doi.org/10.1063/1.5086809

  48. [57]

    Near infrared optical technologies to illuminate the status of the neonatal brain,

    S. M. Liao and J. P. Culver, "Near infrared optical technologies to illuminate the status of the neonatal brain," Curr Pediatr Rev 10 (1), 73 -86 (2014). https://doi.org/10.2174/157339631001140408121507

  49. [58]

    Frequency -domain multiplexing system for in vivo diffuse light measurements of rapid cerebral hemodynamics,

    G. Yu et al., "Frequency -domain multiplexing system for in vivo diffuse light measurements of rapid cerebral hemodynamics," Appl Opt 42 (16), 2931 -2939 (2003). https://doi.org/10.1364/ao.42.002931

  50. [59]

    Frequency domain near -infrared multiwavelength imager design using high-speed, direct analog-to-digital conversion,

    B. B. Zimmermann et al., "Frequency domain near -infrared multiwavelength imager design using high-speed, direct analog-to-digital conversion," J Biomed Opt 21 (1), 16010 (2016). https://doi.org/10.1117/1.JBO.21.1.016010

  51. [60]

    Noninvasive optical monitoring of critical closing pressure and arteriole compliance in human subjects,

    W. B. Baker et al., "Noninvasive optical monitoring of critical closing pressure and arteriole compliance in human subjects," J Cereb Blood Flow Metab 37 (8), 2691-2705 (2017). https://doi.org/10.1177/0271678X17709166

  52. [61]

    Non -invasive estimation of intracranial pressure by diffuse optics: A proof-of-concept study,

    J. B. Fischer et al., "Non -invasive estimation of intracranial pressure by diffuse optics: A proof-of-concept study," J Neurotrauma 37 (23), 2569 -2579 (2020). https://doi.org/10.1089/neu.2019.6965

  53. [62]

    Using near -infrared spectroscopy and a random forest regressor to estimate intracranial pressure,

    F. A. J. Relander et al., "Using near -infrared spectroscopy and a random forest regressor to estimate intracranial pressure," Neurophotonics 9 (4), 045001 (2022). https://doi.org/10.1117/1.NPh.9.4.045001. 27

  54. [63]

    Estimating intracranial pressure using pulsatile cerebral blood flow measured with diffuse correlation spectroscopy,

    A. Ruesch et al., "Estimating intracranial pressure using pulsatile cerebral blood flow measured with diffuse correlation spectroscopy," Biomed Opt Express 11 (3), 1462-1476 (2020). https://doi.org/10.1364/BOE.386612

  55. [64]

    Clinical translation of noninvasive intracranial pressure sensing with diffuse correlation spectroscopy,

    S. Tabassum et al., "Clinical translation of noninvasive intracranial pressure sensing with diffuse correlation spectroscopy," J Neurosurg 139 (1), 184 -193 (2023). https://doi.org/10.3171/2022.9.JNS221203

  56. [65]

    Fast blood flow monitoring in deep tissues with real -time software correlators,

    D. Wang et al., "Fast blood flow monitoring in deep tissues with real -time software correlators," Biomed Opt Express 7 (3), 776 -797 (2016). https://doi.org/10.1364/BOE.7.000776

  57. [66]

    Laser coherence length requirement for speckle contrast diffuse correlation tomography (scdct),

    F. Akbari et al., "Laser coherence length requirement for speckle contrast diffuse correlation tomography (scdct)," SPIE BIOS; Proceedings Volume Multiscale Imaging and Spectroscopy V; PC128270P (2024). https://doi.org/10.1117/12.3001788

  58. [67]

    Choosing a camera and optimizing system parameters for speckle contrast optical spectroscopy,

    T. Y. Cheng et al., "Choosing a camera and optimizing system parameters for speckle contrast optical spectroscopy," Sci Rep . 14 (1), 11915 (2024). https://doi.org/10.1038/s41598-024-62106-y. 28 Caption List Fig. 1: An affordable, wearable, fiber -free PM -DSCF system for deep...

Pith tools

Reviewed August 8, 2026 · model on record in the stance chip above.